4-4
4.5 Replacing consumables
4.5.1 Fuses
■
Disconnect the power plug from the outlet when replacing fuses. Otherwise
you may get electric shock, resulting in death or serious injuries.
■
Use fuses specifically designed for this instrument.
■
When the instrument does not work correctly after fuses have been replaced,
there may be other causes of the problem. Turn off the instrument immediately
and contact your local distributor.
1) Turn power off and disconnect the
power plug from the outlet.
2) Disconnect the power connector from
the power socket of the instrument.
Insert a flat screwdriver or coin into the
slot on the fuse case located at the
bottom of the main unit. Turn the screw
counterclockwise to remove it.
3) Replace the blown fuse with a new
one.
4) Install the fuse holder in the reverse
order of removal.
4.5.2 Built-in printer paper
■
Always use genuine TOMEY paper for the printer. Using other types of paper
may cause printer failure.
■
Do not pull paper forcibly. Trying to pull the paper out may cause printer failure.
Touch the “PRINT”
button without any data stored to feed blank paper.
Replace the paper roll according to the procedure below when red lines
appear on the edges of the printer paper.
1) Push the latch (1) on the printer cover
upward to open it.
(1)
(Fig.1)
(Fig.1)
Содержание REM 4000
Страница 2: ......
Страница 26: ...2 12 This page is intentionally left blank...
Страница 33: ...3 7 Fig 1 Fig 2 2 1 3 4...
Страница 65: ...3 39 2 Enter the ID number and press the save button 2 to change the current examination data ID number Fig 2 2...
Страница 82: ...3 56 This page is intentionally left blank...
Страница 94: ...6 2 This page is intentionally left blank...
Страница 101: ......