Rocket

® 

Quick User Guide

Rocket

® 

PSU

™ 

Portable Suction Unit

If you press 

 you will be asked to 

confirm if you need to change the 
filter.

If you have changed the filter press  

 and you will then be taken to 

the next set up screen. 

If the filter requires replacing press  

 you will be asked to change the 

filter and the unit will be shut down.

Remove the suction unit and change 
the filter as per the instructions, 
re-fit the PSU

 to the chest drain 

bottle and switch on the unit again 
to restart the set up process.

If you press 

 you will progress  

to the next screen.   

Please verify that you have changed 
the filter, by pressing 

 or 

   

to confirm your decision.

If you press 

 you will be taken to 

the next set up screen.

If you press 

 you will be asked 

to change the filter and the unit will 
shut down.

Remove the suction unit and change 
the filter as per the instructions, 
re-fit the PSU

 to the chest drain 

bottle and switch on the unit again 
to restart the set up process.

Содержание PSU Portable Suction Unit

Страница 1: ...Rocket Quick User Guide Rocket PSU Portable Suction Unit...

Страница 2: ...ime 3 Attaching the suction unit to the chest drain bottle 4 5 Operating instructions 6 8 Changing suction levels 9 Information screens 10 11 Switching the unit off 11 Disconnecting the suction unit f...

Страница 3: ...ort is on the back of the device It is protected and is behind the door marked 4 Charge the device for approximately 3hrs until the battery display is full Changing the clock time Please check the tim...

Страница 4: ...r attach tubing set to bottle by inserting straw and cap into aperture and twisting clockwise until clicked shut Take care not to touch the straw 3 Once underwater seal has been established place bott...

Страница 5: ...the PSU over the filter unit so the filter will fit into the filter recess under the suction unit 9 Place the PSU onto the chest drain unit With slight downward pressure rotate the locking ring clockw...

Страница 6: ...perators guide During set up you can confirm or cancel your decisions using the three buttons at the side of the screen 2 Verify if the PSU is connected to a new patient THIS INFORMATION IS IMPORTANT...

Страница 7: ...ilter as per the instructions re fit the PSU to the chest drain bottle and switch on the unit again to restart the set up process If you press you will progress to the next screen Please verify that y...

Страница 8: ...tisfactory The levels of suction applied increase in 0 5KPa 5cmH2O and 3 75mmHg increments There is a gravity drainage setting of 0 25KPa When you are happy with the set level of suction press 6 You w...

Страница 9: ...gether for longer than 2 seconds This will return you to the suction level setting screen Repeat instruction number 5 to change the level of suction required and instruction 6 to confirm When the suct...

Страница 10: ...of use Flow is measured from 10L min Screen 2 30 Minute Flow low The screen shows up to 30 minutes of use Flow is measured from 0 5L min Screen 3 Previous Minutes Flow This screen shows the previous 1...

Страница 11: ...ked to check the filter and then switch off the alarm and screen manually The PSU will continue at the programme suction level Switching the unit off When this has been completed turn the device off b...

Страница 12: ...6988 www rocketmedical com ZDOCK293 Rev 05 02 2021 For all your Rocket PSU Portable Suction Unit service enquiries please contact Service Information Ultrasound Technologies Ltd Tel 44 0 1291 425425...

Отзывы: