Roche Diagnostics
B-166
Operator’s Manual · Version 3.1
7 Configuration
Cedex Bio System
Configuration
System
Item
Values
Comment
Language
Language ID
according to
ISO 639-1.
The language English is always included.
Date Format
dd.mm.yyyy
dd-mm-yyyy
yyyy.mm.dd
yyyy-mm-dd
mm/dd/yyyy
Examples for June 14, 2006:
o
dd.mm.yyyy: 14.06.2006
o
mm/dd/yyyy: 06/14/2006
Time Format
24 Hours
12 Hours
Screen Saver
Wait
0 ... 120 min
Time after which the screen saver feature becomes
active.
0 (zero) means that the screen saver function is off.
The screen saver time is linked to the
Auto Log-off
feature. For example, if you define a Screen Saver
Wait time of 5 minutes, then after 5 minutes of inac-
tivity, the screen saver feature becomes active and the
user is automatically logged off, provided
Auto Log-
off
is on.
Auto Log-off
On
,
Off
If you work with
On
the user is automatically logged
off after a certain number minutes of inactivity on the
instrument. This number is defined in the
Screen
Saver Wait
definition.
Sound Effects
On
,
Off
Acoustic signals when handling instrument items,
e.g. clicking sound when typing. (Does not affect the
acoustic signals for alarms and warnings.)
Volume
0 ... 100
Volume of
any
acoustic signal generated by the
instrument, including alarm and warning signals.
0 (zero) means silent.
Host Server
On
The host interface of the instrument is enabled,
allowing it to communicate with a host.
Off
The host interface of the instrument is disabled.
Communication with a host computer is not possible.
Host definitions have no effect.
TRL Check
Off
The actual results for the lower limit of the measuring
range are displayed.
You cannot change the lower limit of the measuring
range (
Low Limit
). See
On
If you choose
On
, you must, for each test, check the
TRL value (lower limit of the measuring range) in the
relevant application documentation. You may have to
manually adjust it. See
If a measured value is below this TRL threshold, the
actual results are not displayed, instead a value such
as <0.5 is displayed, where 0.5 is an example of the
TRL value defined for this test.
Table B-25
System definitions
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...