Operator’s Manual · Version 3.1
7 Configuration
Cedex Bio System
Applications
Ratio Calculation definitions
19
For applications that use a ratio, perform the following steps:
o
Expand the
Ratio Calculation
entry, select
Coefficient w
, and press
.
o
Define one or two coefficients.
o
Define the formula.
20
Press
to confirm the definitions.
Calculation definitions
21
Expand the
Calculation
entry, select
Factor
, and press
.
22
Change the factor and offset, if correlating two different methods.
23
Choose whether valuation (a reference range) should be employed or not.
If you choose
Reference Range
, you need to define an upper and lower limit and
decide which of them should be taken into account or whether both of them.
24
Define whether the lower limit is used, if required.
25
Define whether the higher limit is used, if required.
26
Change the lower limit value, if required.
27
Change the higher limit value, if required.
28
Define, for each control, whether sample results should be flagged that were
generated using a test whose QC results were marked with a flag.
29
If
TRL Check
is on (
Configuration
>
System
), enter the TRL value (lower limit
of measuring range) as defined in the relevant application documentation.
30
Press
to confirm the definitions.
Result Conversion definitions
31
Expand the
Result Conversion
entry, select
Laboratory Unit
, and press
.
32
Change the laboratory unit if you intend to work with units that are different
from the currently specified units.
33
Type the conversion factor, if required.
This factor is required if lab units were defined.
34
Choose whether values should be displayed in standard or laboratory units.
This definition affects displays on screens as well as printouts.
35
Change the decimal position, if required.
This value corresponds to the number of digits after the decimal point that are
displayed on the screens.
36
Press
to confirm the definitions.
WARNING
Incorrect results due to inappropriate formula
The formula defines how the values of the applications and coefficients are mathemat-
ically combined to generate a result.
It is the responsibility of the user to ensure that the formula is appropriate for the appli-
cation that is being defined.
If one of the initial tests for the ratio calculation did not produce a valid result (the
result would be marked with a flag), the displayed ratio result would be
N/A
.
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...