5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes
89
Line code 11: Reagent Lot Information
Example
12_68354•••••_30/05/2018<LF>
q
o
If no lot information is found or no lot was defined,
“??????????” will be used and transmitted (without quotes).
o
If no expiration date is given for the lot, “00/00/0000”
will be used and transmitted (without quotes).
o
If the lot information is shorter than ten characters, the
string will be filled with spaces on the right side.
Example, N123A will be N123A • • • • •
(•) represents a filler space
Line code 12: Control Lot Information
Example
11_N123A•••••_23/04/2006<LF>
q
o
If no lot information is found or no lot was defined,
“??????????” will be used and transmitted (without
quotes).
o
If no expiration date is given for the lot, “00/00/0000”
will be used and transmitted (without quotes).
o
If the lot information is shorter than ten characters, the
string will be filled with spaces on the right side.
Example, N123A will be N123A • • • • •
(•) represents a filler space
Description
Format
Range
Line code
11
Reagent lot
S10
Reagent expiration date
D1
dd/mm/yyyy
y
Line code 11: Reagent Lot Information
Description
Format
Range
Line code
12
Reagent lot
S10
Reagent expiration date
D1
dd/mm/yyyy
y
Line code 12: Control Lot Information