3
• Patient Glucose Testing
50
Applying a blood sample
Once the meter has checked the test strip, the hourglass
icon disappears and you are prompted to apply a blood
sample.
1
Wait until the flashing drop appears in the display
before applying the blood. The meter beeps.
2
Apply the
drop of blood
to the
front edge
(yellow
dosing area) of the test strip. Blood is pulled into
the test strip by capillary action.
Do
not
apply the blood to the top of the strip. Blood
on top of the test strip is not available for testing
because it will not be pulled into the strip.
Once a sufficient blood sample has been detected, the
meter beeps and the measurement begins.
Patient Test
Patient 123456789
Strip Lot 545794
12:48 pm
09/18/13
When applying the sample, position the meter so that
the test strip port is always higher or on the same level
as the blood drop. This prevents any excess blood
from flowing down the strip and entering the meter.
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...