Patient Glucose Testing •
3
41
3
Patient Glucose Testing
3.1
Information regarding blood glucose testing
Preparing to test
The following requirements must be met before you can
perform a test:
■
The Accu-Chek Inform II test strips are available.
■
The code file for the test strip lot in use must be
stored in the meter (see Chapter 6).
■
An operator ID must be entered (with password, if
required), if the meter is configured for login.
■
Glucose control tests specified in the system con-
figuration must be run and completed successfully
before running patient tests. Information whether
glucose control tests are necessary is displayed in
the
Power Up
screen.
■
If configured, the meter may require downloading
stored data to the data management system within
defined time intervals. If such a download (either
via WLAN or by docking the meter in a base unit)
does not happen within the specified time, the
meter is locked (
Download Lockout
) and cannot be
used for testing.
Contrast
Power Up
QC Due: Immediately
12:48 pm
09/18/13
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...