10
• Maintenance and Care
142
10.5 Unlocking a Download Lockout
When using a data management system for configura-
tion, a
Download Lockout
can be configured. This lockout
prevents a meter from being used for testing, if the meter
data have not been downloaded for a defined period of
time. If a download lockout occurs, the operator may use
a STAT test to temporarily override the lockout. If no STAT
test is configured or left, the meter cannot be used until
the stored data is downloaded via network (either wire-
less or by docking the meter).
In cases of e.g., network failures (which make a download
impossible), or other emergencies, it may be necessary to
unlock a meter without access to the data management
system. Since unlocking is a function of the
Setup Menu
,
it is restricted to persons who know the setup password.
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...