Familiarise yourself
with the Accu-Chek
FastClix lancing device.
Insert the lancet drum,
white end first, into the
lancing device. Push
the lancet drum all the
way in until it clicks
firmly into place.
Get the lancet drum
from the pocket in the
carry case.
NOTE
Do not remove the
lancet drum until all 6
lancets have been used
since the lancet drum
cannot be reused once
it has been removed
from the lancing device.
NOTE
Each lancet drum
contains 6 sterile
lancets.
Replace the cap on
the lancing device.
Make sure the notch
on the cap lines up
with the notch on the
lancing device.
Remove the cap by
pulling it straight off.
Do not twist the cap.
Adjust the lancet
depth to a comfortable
penetration level. Start
at 2 or 3. For tougher
skin, dial to a higher
number.
Using the Accu-Chek
FastClix lancing device:
By pressing the release
button, the lancing
device primes and
pricks in 1-Click
action.
Setting up the Lancing Device
1
4
6
7
5
Accu-Chek
®
Aviva meter
Quick Reference Guide
2
3
WARNING
You must not insert
the lancet drum into
the lancing device and
simultaneously press
the release button or
hold the lancing device
with the release button
resting on a surface
such as a table top.
This could release a
lancet and inadvertently
cause injury.
WARNING
This Quick Start Guide does not replace the User’s
Manual for your Accu-Chek Aviva blood glucose
meter. The User’s Manual contains important handling
instructions and additional information.
WARNING
The Accu-Chek FastClix lancing device is intended for patient
self-monitoring by an individual person only. It must not be
used to collect blood from more than one person as it does not
incorporate any features to guard against cross-infection.
Meter
Overview
Display
Test Strip Slot
Power/Set
Button
Power/Set
Button
Right Arrow
and Left Arrow
Buttons
Battery
Door
Infrared (IR)
Window