Robin AR6004 Скачать руководство пользователя страница 2

 OPERATION

Before taking any measurements, read the Safety Infor-

mation Section. Always examine the instrument for dam-

age, contamination (excessive dirt, grease, etc.) and de-

fects. Examine the test leads for cracked or frayed insula-

tion. If any abnormal conditions exist do not attempt to

make any measurements.

Auto Power-down mode

If unused for about 30 minutes, the tester will power-

down automatically. Press RESET button  to resume

power-on mode. If you should disable Auto Power-down

mode, press SELECT button when you turn on the meter.

SELECT Function Button (

/

)

The SELECT Function button is Yellow  in color. Press

it to toggle to the alternate function ( Audible continuity and

Diode ) shown in Yellow on the meter face.

RESET Button

Press RESET button to reset or initialize system logic.

If you suspect that the meter is not operating properly or the

meter was power-down, you can always restart from the

beginning by press the RESET button.

RANGE Button

Press (RANGE) button to select the Manual Range mode

and turn off the "AUTO" annunciator.

In the Manual Range mode. Each time you press

(RANGE) button, the range (and the input range annuncia-

tor) increments, and a new value is displayed. To exit the

Manual Range mode and return to autoranging, press and

hold down (RANGE) button for 2 seconds. The "AUTO"

annunciator turns back on.

REL 

∆∆∆∆∆

 Button

Press (REL 

) to enter the Relative mode, zero the

display, and store the displayed reading as a reference

value. The relative mode annunciator (

) is displayed.

Press (REL 

) again to exit the relative mode.

Data-Hold Button

Press this button briefly to activate DATA-HOLD mode.

The " 

H

 " annunciator is displayed.

Press this button again to exit the DATA-HOLD mode.

Voltage Measurements

1. Connect the red test lead to the "V

" jack and the black

test lead to the "COM" jack.

2. Set the Function/Range switch to the desired voltage

function.

3. Connect the test leads to the device or circuit being

measured.

4. For dc, a (-) sign is displayed for negative polarity;

positive polarity is implied.

Resistance Measurements

1. Set the Function/Range switch to 

/

 position.

2. Remove power from the equipment under test.

3. Connect the red test lead to the "V

" jack and the black

test lead to the "COM" jack.

4. Touch the probes to the test points. In ohms, the value

indicated in the display is the measured value of resistance.

WARNING

The accuracy of the functions might be slightly affected,

when exposed to a radiated electromagnetic field environ-

ment, e.g., radio, telephone or similar.

Diode Tests

1. Set the Function/Range switch to 

/

 position.

2. Remove power from the equipment under test.

3. To toggle the 

/continity/diode modes, press SELECT

switch.

4. Touch probes to the diode. A forward-voltage drop is

about 0.6V (typical for a silicon diode).

5. Reverse probes. If the diode is good, 1 OL is displayed.

If the diode is shorted, ".000" or another number is

displayed.

6. If the diode is open, 1 OL is displayed, in both directions.

7. If the junction is measured in a circuit and a low reading

is obtained with both lead connections, the junction may

be shunted by a resistance of less than 1k

Ω. 

In this case

the diode must be disconnected from the circuit for

accurate testing.

EMC

Errors are 

10% at field strengths below 1V/m and will

increase at higher field strengths.

GROUND

DOUBLE INSULATION
(Protection ClassII)

DC-DIRECT
CURRENT

AC-ALTERNATING
CURRENT

SEE EXPLANATION
IN MANUAL

DANGEROUS
VOLTAGE

International Electrical Symbols

MAINTENANCE

WARNING

Remove test leads before changing battery or fuse or

performing any servicing.

Battery Replacement

Power is supplied by a 9 volt "transistor" battery.

(NEDA 1604, IEC 6F22). The " " appears on the LCD

display when replacement is needed. To replace the bat-

tery, remove the two screws from the back of the meter and

lift off the battery case. Remove the battery from battery

contacts.

Servicing

Refer all service and repair to Robin Electronics. Do not

attempt servicing or repair yourself. Robin Electronics

recommend this instrument is calibrated at least once every

12 months. Contact Robin for a competitive quotation.

Cleaning

Periodically wipe the case with a damp cloth and

detergent, do not use abrasives or solvents.

Continuity Measurements

1. Set the Function/Range switch to 

/

 position.

2. Turn off power to the circuit under test. External voltage

across the components causes invalid readings.

3. To toggle between the 

/continity/diode modes, press

SELECT button.

4. Connect the test leads to the two points at which

continuity is to be tested. The buzzer will sound if the

resistance is less than approximately 100

Ω.