RND lab 355-00011 Скачать руководство пользователя страница 4

6

7

EN

EN

RESISTANCE [ Ω ] MEASUREMENT

WARNING: To avoid electric shock, disconnect power to the unit under test and discharge 

all capacitors before taking any resistance measurements. Remove the batteries and 

unplug the line cords.

1. Set the function switch to the Ω position.

2. Insert the black test lead plug into the negative (COM) socket and the red test lead plug

into the positive Ω jack.

3. Press the MODE button until “Ω” appears in the display.

4. Touch the test probe tips across the circuit or part under test. It is best to disconnect

one side of the part under test so the rest of the circuit will not interfere with the

resistance reading.

5. Read the resistance in the display. The display will indicate the proper decimal point,

value and symbol.

CONTINUITY CHECK

WARNING: To avoid electric shock, never measure continuity on circuits or wires that have 

voltage on them.

1. Set the range switch to the •))) position.

2. Insert the black lead plug into the COM socket and the

red test lead plug into the positive •))) socket.

3. Press the MODE button until “•)))” appears in the display.

4. Touch the test probe tips to the circuit or wire you wish to check.

5. If the resistance is less than 35Ω, the audible signal will sound. The display will also

show the actual resistance in ohms.

DIODE TEST

WARNING: To avoid electric shock, do not test any diode that has voltage on it.

1. Set the function switch to the          position.

2. Insert the black test lead plug into the COM socket and the red test lead plug into the

             socket.

3. Press the MODE button until “          ” appears in the display.

4. Touch the test probe tips to the diode or semiconductor junction you wish to test. Note

the meter reading.

5. Reverse the probe polarity by switching probe position. Note this reading.

6. The diode or junction can be evaluated as follows:

A. If one reading shows a value and the other reading shows OL, the diode is good.

B. If both readings show OL, the device is open.

C. If both readings are very small or zero, the device is shorted.

NOTE: The value indicated in the display during the diode check is the forward voltage.

CAPACITANCE MEASUREMENT

WARNING: To avoid electric shock, discharge the capacitor under test before measuring.

1. Set the function switch to the CAP capacitance position. Press the MODE button until

“nF” appears in the display.

2. Insert the black test lead banana plug into the negative COM jack and the red test lead

banana plug into the CAP positive jack.

3. Touch the test probe tips across the part under test.

4. Read the capacitance value in the display.

5. The display will indicate the proper decimal point and value.

Note: For very large values of capacitance measurement time can be several minutes 

before the fi nal reading stabilizes. The bar graph is disabled in capacitance measurement 

mode. The LCD displays DIS. C .Discharging through the chip is quite slow. We 

recommend the user to discharge the capacitor with some other apparatus.

FREQUENCY MEASUREMENT

1. Set the function switch to the Hz position.

2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead

banana plug into the  positive Hz jack.

3. Touch the test probe tips to the circuit under test.

4. Read the frequency in the display. The digital reading will indicate the proper decimal

    point, symbols ( Hz ,kHz, MHz) and value.

TEMPERATURE MEASUREMENT

1. Set the function switch to the Temperature position.

2. Insert the Temperature Probe into the input jacks, making sure to observe the correct

    polarity.

3. Press the “°F/°C” button until “°F or °C” appears in the display.

4. Touch the Temperature Probe head to the part whose

temperature you wish to measure. Keep the probe touching the part under test until the

reading stabilizes (about 30 seconds).

5. Read the temperature in the display.

Note: The temperature probe is fi tted with a type K mini connector. A mini connector to 

banana connector adaptor is supplied for connection to the input banana jacks.

Содержание 355-00011

Страница 1: ...www rnd electronics com Autoranging Digital Multimeter User Manual RND 355 00011...

Страница 2: ...ct This symbol advises the user that the terminal s so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds in this case 1000 VAC or VDC Equipment...

Страница 3: ...ition 3 Use the MODE button to select AC or DC Voltage 4 Connect the test leads in parallel to the circuit under test 5 Read the voltage measurement on the LCD display DC CURRENT MEASUREMENT 1 Insert...

Страница 4: ...on can be evaluated as follows A If one reading shows a value and the other reading shows OL the diode is good B If both readings show OL the device is open C If both readings are very small or zero t...

Страница 5: ...ication Low battery indication A battery symbol is displayed when the battery voltage drops below the operating level Measurement rate 2 times per second nominal Auto power off Meter automatically shu...

Страница 6: ...range 10A dc on 10A range AC Current Auto ranging Overload Protection FF500mA 1000V and F10A 1000V Fuse AC Response 50 Hz to 400 Hz Maximum Input 400uA ac rms on uA 400mA ac rms on mA 10A ac rms on 1...

Страница 7: ...below the operating voltage the battery warning symbol will appear in the LCD display The battery should be replaced 2 Follow instructions for installing battery See the Battery Installation section o...

Страница 8: ...om its holder by gently pulling it out 4 Install the new fuse into the holder 5 Always use a fuse of the proper size and value 0 5A 1000V fast blow for the 400mA range 10A 1000V fast blow for the 10A...

Страница 9: ...nleitung beziehen muss um Verletzungen oder Sch den am Messger t zu vermeiden Achtung Stromschlaggefahr Dieses WARNUNG Symbol weist auf eine potenziell gef hrliche Situation hin die zu schweren oder t...

Страница 10: ...pannung aus 4 Schliessen Sie die Messleitungen parallel an den zu pr fenden Stromkreis an 5 Lesen Sie die Spannungsmessung auf der LCD Anzeige ab DC STROMMESSUNG 1 Stecken Sie den schwarzen Bananenste...

Страница 11: ...er Messung ein Wert und bei der anderen OL angezeigt wird ist die Dio de in Ordnung B Wenn bei beiden Messungen OL angezeigt wird ist das Ger t defekt sperrt in beide Richtungen C Wenn die beiden Mess...

Страница 12: ...Das Messger t schaltet sich nach ca 30 Minuten Inaktivit t automatisch ab Betriebsumgebung 10 C bis 50 C 14 F bis 122 F bei 70 relativer Luftfeuchte Lagertemperatur 30 C bis 60 C 4 F bis 140 F bei 80...

Страница 13: ...0 des Messwerts 5 Stellen berlastschutz FF500 mA 1000 V und F10 A 1000 V Sicherung AC Reaktion 50 Hz bis 400 Hz Maximaler Eingang 400 A AC eff auf A 400 mA AC eff auf mA 10 A AC eff im 10 A Bereich Wi...

Страница 14: ...ICHERUNG austauschen WARNUNG Um einen Stromschlag zu vermeiden trennen Sie die Messleitungen von einer Spannungsquelle bevor Sie die Batteriefachabdeckung entfernen 1 Wenn die Batterien leer sind oder...

Страница 15: ...endreher l sen 3 Entfernen Sie die alte Sicherung aus ihrer Halterung indem Sie sie vorsichtig herausziehen 4 Setzen Sie die neue Sicherung in den Halter ein 5 Verwenden Sie immer eine Sicherung der r...

Страница 16: ...3 DE DE...

Страница 17: ...www rnd electronics com...

Отзывы: