5 Operation
Manual GC 9300 EN14 July, 24th 2018
51
V.
Click the "Apply" button.
VI.
Click the "Back" button to exit the parameterization window.
5.2.2
Change operating mode
To change the operating mode, you must first enter the user code. Stay on the
"Details" screen and proceed as follows:
I.
Scroll up to the top of the window on the left.
II.
Click on "GC9300 Mode".
III.
Click on "Operating Mode" in the window on the right.
IV.
Select the desired operating mode in the selection field.
V.
Click the "Apply" button.
VI.
Click the "Back" button to exit the parameterization window.
The following operating modes are available:
AUTORUN
Normal analysis operation. One analysis is performed after
another, interrupted by automatic calibration.
STOP
Measurement and calibration operation are disabled.
BASIC-CAL
Basic calibration for startup and for RMG service. Basic cali-
bration may not be carried out during normal analysis opera-
tion.
NORM-CAL
Manual calibration, which can be started at any time and has
the same effect as automatic calibration and also takes 30
minutes. AUTORUN is activated automatically after calibra-
tion.
REF-GAS
Reference gas analysis, i.e. the reference input for the analy-
sis of test gas or a gas sample is connected. The PGC re-
mains in this mode until it is changed again or the maximimum
of reference gas analysis are done.
After the operating mode is changed, the analysis or calibration currently being
performed is ended and switchover takes place directly afterwards.