65-2483RK Multi Point Detector • 31

Calibration, O

2

 Detector

This section describes how to calibrate the oxygen detector on the multi-point detector. It 
includes procedures to prepare for calibration, set the fresh air reading, set the zero 
reading, and return to normal operation. It describes calibration using a calibration kit 
that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator 
with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per 
minute) fixed flow regulator. To properly calibrate the multi-point detector, you must 
calibrate each connected detector. Perform the following procedure for the oxygen 
detector.

WARNING: The controller is not an active gas monitoring device during the calibration 

procedure.

Preparing for Calibration

WARNING: Do not remove the junction box cover or detector housing cap while the 

circuits are energized unless the area is determined to be non-hazardous. 
Keep the junction box cover and detector housing cap tightly closed during 
operation.

1.

Screw the calibration cup onto the bottom of the oxygen detector.

2.

Screw the regulator into the zero air calibration cylinder. Make sure the regulator is 
off. It is off when the on/off knob is turned all the way clockwise.

NOTE

If you can verify that the multi-point detector is in a fresh air environment, you 
do not need to apply zero air to the detector before adjusting the fresh air 
reading.

3.

Use the calibration kit sample tubing to connect the fixed flow regulator to the 
calibration cup.

4.

Put the controller into its calibration program. See the controller operator’s manual 
for instructions to enter the calibration program.

NOTE:

RKI controllers have a peak hold feature for fresh air adjustment and a minimum 
hold feature for zero adjustment. Because of this, the instructions call for turning 
off the gas to a detector before making adjustments.

Setting the Fresh Air Reading

1.

Follow the directions in the controller operator’s manual for setting the fresh air 
reading.

2.

When the instructions call for applying zero air to the detector, turn the regulator’s 
on/off knob counterclockwise to open it. Gas will begin to flow.

3.

Allow the gas to flow for two minutes.

4.

Turn the regulator’s on/off knob clockwise to close it.

5.

Set the fresh air reading according to the controller operator’s manual.

6.

Unscrew the regulator from the zero air calibration cylinder. Leave the sample tubing 

Содержание 65-2483RK

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2483RK Multi Point Detector Operator s Manual Part Number 71 0206RK Revision A Released 8 2 11...

Страница 2: ...calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type...

Страница 3: ...ent costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPR...

Страница 4: ...he Zero Fresh Air Reading 20 Maintenance 21 Preventive Maintenance 21 Troubleshooting 23 Replacing Components of the Multi Point Detector 24 Calibration LEL H2S and CO Detectors 29 Preparing for Calib...

Страница 5: ...RK 03 LEL IR HC H2S 65 2483RK 04 LEL IR HC CO 65 2483RK 05 LEL IR HC O2 65 2483RK 06 LEL IR HC H2S CO Note The standard factory calibration for the IR HC LEL detector is to propane Since the 65 2483RK...

Страница 6: ...tation for installation and operation instructions WARNING When using the 65 2483RK you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65 2483RK an...

Страница 7: ...rared combustible gas LEL HC detector housed and encapsulated in a pipe nipple The pipe nipple has 3 4 NPT threads on each end and a 1 1 4 hex that allows removal or installation of the detector with...

Страница 8: ...s the sensor s signal before the signal reaches the controller Housing Cap Cap Gasket The housing cap screws onto the detector housing It retains the plug in sensor and protects it from damage A foam...

Страница 9: ...p of the detector housing body Use these wires to connect the CO detector to the multi point detector The housing includes a four socket pattern This socket pattern accepts the sensor s four pins to s...

Страница 10: ...ern This socket pattern accepts the sensor s four pins to secure the sensor within the detector housing A pre amplifier located between the sockets and two interconnect wires conditions the sensor s s...

Страница 11: ...e Select a site where the detector is not likely to be bumped or disturbed Make sure there is sufficient room to perform start up maintenance and calibration procedures Select a site where the target...

Страница 12: ...ti Point Detector Figure 6 Outline and Mounting Dimensions 65 2483RK 02 3 4 NPT HUBS 4X 3 11 25 Oxygen Detector 1 10 2 40 CO Detector LEL HC IR Detector 3 00 3 15 7 75 3 00 1 25 20 DIA X 45 Slot 2X mo...

Страница 13: ...Point Detector 9 Figure 7 Outline and Mounting Dimensions 65 2483RK 03 3 11 3 4 NPT Hubs 4X 25 LEL HC IR Detector 1 10 2 40 H2S Detector 3 4 Conduit Plug 3 00 3 15 7 63 3 00 1 25 20 DIA X 45 Slot 2X m...

Страница 14: ...ti Point Detector Figure 8 Outline and Mounting Dimensions 65 2483RK 04 25 3 11 3 4 NPT Hubs 4X LEL HC IR Detector 1 10 2 40 CO Detector 3 4 Conduit Plug 3 00 3 15 7 63 20 DIA X 45 Slot 2X mounting 1...

Страница 15: ...int Detector 11 Figure 9 Outline and Mounting Dimensions 65 2483RK 05 3 4 NPT HUBS 4X 25 3 11 LEL HC IR Detector 1 10 20 DIA X 45 Slot 2X mounting 2 40 Oxygen Detector 3 00 3 4 Conduit Plug 3 15 7 63...

Страница 16: ...ng site you select use 10 screws to mount the detector to a vertical surface CAUTION Mount the multi point detector with the detectors facing down see Figures 5 10 3 4 NPT HUBS 4X 3 11 25 1 10 CO Dete...

Страница 17: ...The junction box is rated NEMA 4X and classified explosion proof for Class I Groups B C and D 5 Connect the wires to the multi point detector using the terminal block in the junction box CAUTION If u...

Страница 18: ...al Strip Oxygen Detector Green Controller H2S Terminals Controller LEL Terminals Controller Oxygen Terminals Black Red Red White Green White Black Green LEL HC IR Detector W ires Oxygen Detector Wires...

Страница 19: ...nal Strip Green Oxygen Detector Green Red Controller CO Terminals Controller LEL Terminals Controller Oxygen Terminals Black Red White Green White Black LEL HC IR Detector W ires Oxygen Detector Wires...

Страница 20: ...iring the 65 2483RK 03 to a Controller Terminal Strip Black LEL HC IR Detector Controller LEL Terminals Controller H2S Terminals Black Red Red White Black Green LEL HC IR Detector W ires Green White R...

Страница 21: ...Wiring the 65 2483RK 04 to a Controller Red Black Terminal Strip Green Black LEL HC IR Detector Controller CO Terminals Controller LEL Terminals White Red Red White LEL HC IR Detector W ires Green Bl...

Страница 22: ...g the 65 2483RK 05 to a Controller White Terminal Strip LEL HC IR Detector Wires LEL HC IR Detector Controller LEL Terminals Controller OXY Terminals Red White Green Green Black Red White Green Black...

Страница 23: ...ontrollers typically have a ground stud that can be used to ground the cable s drain wire Red White Black Green Red Black Controller H2S Terminals Controller LEL Terminals Controller CO Terminals CO D...

Страница 24: ...the area is determined to be non hazardous Keep the detector housing cap and junction box cover tightly closed during operation CAUTION If you suspect the presence of a target gas in the monitoring e...

Страница 25: ...ed flow regulator NOTE Performing a response test on the multi point detector may cause alarms Be sure to put the controller into its calibration program or disable external alarms before performing t...

Страница 26: ...ect response reading calibrate the detector as described in the Calibration section of this manual 3 Turn the regulator s on off knob clockwise to close the regulator 4 Unscrew the regulator from the...

Страница 27: ...set the zero or response reading during calibration Detector requires frequent calibration Note Under normal circumstances the transmitter requires calibration once every 3 months for O2 H2S and CO d...

Страница 28: ...e IR LEL HC detector to verify the connections are correct Also see Figure 11 Figure 16 for wiring diagrams 8 Reinstall the junction box cover 9 Turn on or plug in power to the controller 10 Turn on t...

Страница 29: ...oded leads as you remove them 5 Unscrew the detector from the junction box hub 6 Guide the detector leads of the replacement detector through the junction box hub then screw the mounting threads of th...

Страница 30: ...ated in the detector housing 8 Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body 9 Turn on or plug in power to the controller 10 Turn on the c...

Страница 31: ...fy the connections are correct Also see Figure 11 Figure 16 for wiring diagrams 8 Reinstall the junction box cover 9 Turn on or plug in power to the controller 10 Turn on the controller and place into...

Страница 32: ...the detector into the hub If necessary for environmental conditions apply thread sealant or teflon tape to the hub and or detector threads to seal them 7 Connect the detector leads to the terminal bl...

Страница 33: ...To properly calibrate the multi point detector you must calibrate each connected detector Perform the following procedure for each connected LEL H2S or CO detector WARNING The controller is not an act...

Страница 34: ...e channel you are calibrating 2 Follow the directions in the controller operator s manual for setting the response reading span 3 When the directions call for exposing the detector to gas turn the reg...

Страница 35: ...egulator into the zero air calibration cylinder Make sure the regulator is off It is off when the on off knob is turned all the way clockwise NOTE If you can verify that the multi point detector is in...

Страница 36: ...ve regulator and calibration cup connected by the sample tubing 2 When the controller display reading rises above the alarm points return the controller to normal operation NOTE If you do not allow th...

Страница 37: ...el 81 0076RK Zero air calibration cylinder 17 liter steel 81 0076RK 01 Zero air calibration cylinder 34 liter steel 81 0076RK 03 Zero air calibration cylinder 103 liter steel 81 0078RK 01 Calibration...

Отзывы: