16 • 65-2443SS-XX PPM Hydrogen Transmitter

11. Allow the transmitter to run for 3 hours.
12. Perform a preliminary calibration as described on page 16. This will be a temporary calibration 

since the new ppm detector must burn-in before it can be properly calibrated.

13. Allow the replacement detector to burn-in for at least 5 days

 

before you continue with the next 

step. A burn-in period of 7 days is recommended.

14. After the burn-in period, calibrate the transmitter as described on page 16.

NOTE:

The J3 and J4 jumper configuration may need to be adjusted during calibration.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all ppm hydrogen transmitter 
applications, a calibration frequency of every 6 to 12 months is adequate for most applications. Unless 
experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a 
calibration frequency of every 6 months. 

If an application is not very demanding, for example detection in a clean, temperature controlled 
environment where hydrogen is not normally present, and calibration adjustments are minimal at 
calibration, then a calibration frequency of every 12 months is adequate. 

If an application is very demanding, for example if hydrogen is present often and in significant 
concentrations or the environment is not well controlled, then more frequent calibration than every 6 
months may be necessary.

Calibration

This section describes how to calibrate the ppm hydrogen transmitter. It includes procedures to prepare 
for calibration, set the span reading, set the linearization, and return to normal operation.

WARNING: The controller is not an active gas monitoring device during the calibration 

procedure.

Calibration Kit Humidifier Tube

The ppm hydrogen detector requires normal atmospheric humidity levels to respond properly to 
hydrogen. Normal atmospheric humidity variations do not affect the detector’s response to hydrogen 
in ambient air, but the ultra low humidity level of gas from a calibration cylinder requires that the 
calibration sample be humidified for the detector to respond properly. The calibration kit for the ppm 
hydrogen transmitter includes a humidifier tube that is not normally included in other calibration kits. 
This humidifier tube humidifies the calibration sample flowing through it by absorbing humidity from 
the ambient air and adding it to the sample. The humidifier tube is included in the “Parts List” on 
page 19.

WARNING: A humidifier tube must be used when calibrating the ppm hydrogen transmitter for 

the detector to respond properly to the calibration gas. Failure to use a humidifier 
tube will result in an inaccurate calibration.

Содержание 65-2443SS Series

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2443SS XX PPM Hydrogen Transmitter Operator s Manual Part Number 71 0549 Revision P2 Released 3 19 21...

Страница 2: ...odic calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the...

Страница 3: ...epair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRESSED OR IMPLIED...

Страница 4: ...artup 12 Introducing Incoming Power 12 Transmitter Burn In Period 12 Setting the Zero Signal 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 13 Replacing Components of the PPM Hydrogen Tra...

Страница 5: ...he 65 2443SS XX you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65 2443SS XX and to minimize the risk of personal injury Be sure to maintain and...

Страница 6: ...n the detector The 1 2 NPT mounting threads at the top of the detector allow you to mount it into the bottom conduit hub of the junction box A rainshield screws onto the bottom of the detector flame a...

Страница 7: ...ector terminal strip to connect the detector to the amplifier NOTE The detector is factory wired to the detector terminal strip See page 10 for all wiring procedures related to the transmitter Span Po...

Страница 8: ...on one pin When installing a new sensor the jumper configuration may need to be adjusted see page 16 Junction Box The stainless steel corrosion resistant junction box allows you to install the detect...

Страница 9: ...mounting site Select a site where the transmitter is not likely to be bumped or disturbed Make sure there is sufficient room to perform start up maintenance and calibration procedures Select a site wh...

Страница 10: ...through the top conduit hub of the junction box 7 Pull out the controller terminal strip and connect the three wires to the terminal strip as follows see Figure 4 Connect the positive wire to the PWR...

Страница 11: ...PPM Hydrogen Transmitter to a Controller 12 If shielded cable is used connect the cable s drain wire to an available chassis earth ground at the controller RKI controllers typically have a ground stud...

Страница 12: ...n on the controller 5 Verify that the controller is on and operating properly Refer to the controller operator s manual Transmitter Burn In Period Once power has been applied to the transmitter it wil...

Страница 13: ...Probable Causes Recommended Action Fail Condition Controller indicates a fail condition The transmitter wiring is disconnected or misconnected The transmitter is malfunctioning 1 Verify that the tran...

Страница 14: ...install the detector terminal strip into its socket NOTE When a transmitter is first powered up with a new detector its output may be high enough to cause an alarm or low enough to indicate a failure...

Страница 15: ...n Figure 4 on page 11 of this manual If controller leads or detector leads were removed during this procedure refer to Tables 4 and 5 below 8 Install the amplifier into the junction box with the screw...

Страница 16: ...ple if hydrogen is present often and in significant concentrations or the environment is not well controlled then more frequent calibration than every 6 months may be necessary Calibration This sectio...

Страница 17: ...cover then plug the voltmeter leads into the test points on the amplifier Plug the positive lead into the red test point plug the negative lead into the black test point 5 Place the controller into i...

Страница 18: ...V 2 mV on the voltmeter install the J3 jumper see page 7 and try to adjust the span pot again 5 Turn the regulator knob clockwise to close the regulator 6 Unscrew the regulator from the full scale cal...

Страница 19: ...000RK 21 Calibration cylinder 2 000 ppm hydrogen in air 34 liter steel 81 0000RK 23 Calibration cylinder 2 000 ppm hydrogen in air 103 liter 81 0000RK 71 Calibration cylinder 500 ppm hydrogen in air 3...

Страница 20: ...0 ppm hydrogen steel calibration cylinder and 34 liter 1 000 ppm hydrogen steel calibration cylinder 81 F074RK Calibration kit includes regulator humidifier tube calibration cup 103 liter 250 ppm hydr...

Отзывы: