Setting Detailed Operational Parameters Using the SDM-PC2 Docking Station PC
SDM-6000 within this time period, “Cal Limit Soon” will display in the text field
above the BUMP and CAL buttons in the Function Tab.
5 . Select the desired setting for the
Cylinder Expiration Notice
parameter. This
is the length of time, in days, before a cylinder’s expiration date that the
SDM-6000 will give an indication that the cylinder’s expiration date is
approaching. So if the cylinder expires on 12/20/15 and the
Cylinder
Expiration Notice
is set to
10
days, the cylinder-expiring indication will begin
appearing on 12/10/15. The cylinder icon in the bottom left corner of the
SDM-6000’s Function Tab will change from green to orange.
6 . To save the changes, click
SDM Update
and click
Yes
in the window that
appears.
To return to the main window without saving any changes, click
Cancel
and
click
OK
in the window that appears.
Updating the Automatic Execution Settings
The Automatic Tab allows you to set up a bump test, calibration, or alarm check to
automatically occur at a specific time on the desired day(s). If a GX-6000 is installed
in an SDM-6000 at the specified day/time, the SDM-6000 will turn the GX-6000 on,
perform the test, then turn the GX-6000 off.
1 . Launch the PC Controller Program and connect your SDM-6000(s) as
described in “Launching the PC Controller Program and Connecting an SDM-
6000” on page 60.
2 . Access the SDM Configuration Window as described in “Accessing the SDM