SD-1GH
- 32 -
CAUTION
A reading above zero is suppressed with the 10% FS suppression.
A reading that is 10% FS or more below zero is displayed as "-0.0". The detector needs a zero
adjustment and span adjustment.
6-4. Suppression Function
The sensors used with the detector are influenced by environmental changes (temperature, humidity, and
other characteristics) and interference gases, which affects the reading. Therefore, the reading might
fluctuate around zero even in a normal environment.
The suppression function obscures influences by environmental changes and interference gases around
zero that have no meaning for your management rules for a gas alarm. This function is used to hide
(suppress) the fluctuation of the reading around zero.
NOTE
•
The suppression function is factory-set. The standard setting value is 10% FS.
•
In Maintenance Mode, this function is disabled and the fluctuation around zero is displayed.
<<Example>>
Suppression Disabled
A fluctuation around zero is displayed as the reading.
Suppression Enabled
A fluctuation around zero is displayed as zero.
(Zero suppression)