30
•
Chapter 3: Installation & Startup
M2A-XL Transmitter Operator’s Manual
Figure 20: Outline & Mounting Dimensions, CT-7 Series
Wiring the M2A-XL Transmitter
WARNING: Always verify that the power source is OFF before making any wiring
connections.
1. Remove the junction box cover.
2. Grasp the control PCB by its edges.
3. Gently pull until the control PCB is pulled away from the banana jacks. Take care not to pull
too hard and damage the cable which connects the control and terminal PCBs.
4. Let the control PCB hang by the cable. The terminal strips are now visible on the terminal
PCB. The control PCB may be left hanging while wiring is done. If desired, the control PCB
may be disconnected from the cable and set aside while wiring.
5. The detector leads are factory wired. Verify that the detector leads are wired to the detector
terminal strip as shown in the applicable figure below:
3/4 NPT
Conduit Port,
Plugged
1.40 max
1.00
4.86
3/4 NPT
Conduit Port
Ø 0.25, 2X
5.5
5.66
2.82
6.2
6.90 max