- 46 -
7-3. Span Adjustment
<Simultaneous Span Adjustment for All Channels (AUTO CAL)>
AUTO CAL
Press the ENTER switch.
The span gas concentration is displayed.
Use the
or
switch to select the standard 4
sensors or the high concentration combustible gas
sensor (if installed).
The span gas concentration(s) shown need to be
consistent with the concentration of the prepared
span gas. To change the value, press the
and
DISPLAY switches simultaneously.
The span gas concentration setting menu is
displayed.
Press the
or
switch to display a target gas for
which the span gas concentration should be
changed. Every time AIR is pressed, the target gas
is changed (in the order of Combustible, O2, H2S,
CO, and ESCAPE).
While a desired target gas is selected, press the
ENTER switch to cause the span gas
concentration display to blink, prompting for input.
Press the
or
switch to change the span gas
concentration setting.
After the ENTER switch is pressed to confirm the
changed setting and END is displayed, the gas
monitor automatically returns to the span gas
concentration setting menu. Repeat the same
procedure to set other target gases.
When all of the span gas concentration setting is
completed, press the
▲
or
▼
switch until
ESCAPE is displayed. Press ENTER to return to
the original screen.
While the span gas concentration is displayed,
press the ENTER switch to cause AUTO CAL to
blink and display the current concentration
readings (prompting for supply).
Screw the demand flow regulator into the cylinder.
NOTE: Calibration gas for the high concentration
combustible gas sensor MUST be in a balance of
N2 and cannot be in a balance of air.
Use the cal. tubing to connect the regulator to the
probe.
In 60 seconds after supplying the span gas when
the readings are stabilized, press the ENTER
switch to perform adjustment.
If the adjustment is successfully completed, PASS is
displayed, and the gas monitor automatically returns
to the menu.
If the adjustment failed on one or more channels a
screen indicating which channel(s) failed will appear.
Press and release the
▼
switch to return to the
menu.
Disconnect the calibration tubing from the probe.
Return to AIR CAL.
<Span Gas Concentration
Setting>
[Setting repeated for each gas]
NOTE
The changed value is saved.
From next time, the changed
value is displayed in the
beginning.
Supply the span
gas.
Stop supplying
the span gas.