GX-2009 Operator’s Manual
User Setup Mode • 51
4. If the PASSWORD menu item is set to
on
, the following screen will appear with the first
digit flashing prompting you to enter the password.
•
Use the AIR button to scroll through the numbers 0-9 until you arrive at the desired
value, then press and release the POWER MODE button to accept the value.
•
Repeat the above procedure for each digit.
•
When you have set and accepted the last digit of the password, the first menu item in
User Setup Mode will appear, the DATE Screen.
•
If you entered the incorrect password, an error screen will appear.You must use the
POWER MODE button to turn the unit off and try again if you wish to enter User
Setup Mode or turn on the unit and proceed to Measuring Mode.
5. If the PASSWORD menu item is set to
oFF
, the first menu item in User Setup Mode
displays, the DATE Screen.
6. Use the AIR button to move forward through User Setup Mode. When you get to the last
menu item, the START menu item, continuing will take you to the beginning of the menu.
7. When you arrive at the item you wish to enter, press and release the POWER MODE
button to enter that item.
8. When you need to adjust the numerical value of a parameter, increase it or decrease it, use
the AIR button to change the value. When adjusting a numerical parameter value, it is
possible to reverse the direction of adjustment. To switch from increasing to decreasing a
value or decreasing to increasing a value, do the following:
•
with the parameter flashing on the screen, press and hold the AIR button
•
immediately press the POWER MODE button and then release both buttons
•
the direction of adjustment when you press the AIR button is now reversed
0 0 0 0
PASSWORD
DATE
DATE
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30