60 • Chapter 5: User Mode and Calibration
04 Series for CO(-H2)
LED+BUZ
: The CO-04C’s LED double flashes and the buzzer double beeps as often as
defined by the
BEEP.INT
parameter to verify that the instrument is operating.
BMP/CAL
: If a bump test or a calibration is due and if
BP.EXPD
or
CAL.EXPD
is set to
CONFIRM
(factory setting) or
NONE
, the CO-04C’s LED double flashes as often as defined
by the
BEEP.INT
parameter to indicate a non-compliance. Once a bump test or calibration
(depending on which is due) is done, the LED stop flashing.
1. While in the
BEEP
menu, press AIR to scroll to
BEEP.SEL
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BEEP.SEL
item in
the
BEEP
menu.
5. See “ESCAPE” on page 60 to return to User Mode.
BEEP.INT
The
BEEP.INT
parameter defines how often the confirmation alert or non-compliance
indicator selected in
BEEP.SEL
occurs. This setting only applies if the
BEEP.SEL
parameter
is set to something other than
OFF
(factory setting). The available choices are
0.5
minutes
and
1
-
99
minutes in 1 minute increments. The factory setting is
5
minutes.
1. While in the
BEEP
menu, press AIR to scroll to
BEEP.INT
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BEEP.INT
item in
the
BEEP
menu.
5. See “ESCAPE” on page 60 to return to User Mode.
ESCAPE
1. While in the
BEEP
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
BEEP
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 64 to enter Measuring Mode.
BEEP .SEL
BEEP .INT