VOC Pro Transmitter • 57
6. Screw the sensor housing cap back onto the sensor housing base, ensuring that the sensor
housing cap is only tightened hand tight.
7. Zero and calibrate the detector as described on page 59.
Fuse Replacement
WARNING: Do not remove the sensor housing cap or enclosure lid while the circuits are
energized unless the area is determined to be non-hazardous. Keep the sensor
housing cap and enclosure lid tightly closed during operation.
CAUTION:
The internal components can be static sensitive. Use caution when opening the
enclosure and handling internal components. DO NOT use any metal objects or
tools to remove the fuses.
1. Disconnect or turn off power to the VOC Pro.
2. Unscrew the enclosure lid and set it aside.
3. Grab the thumbscrews and gently lift the internal system out of the enclosure. It can rest
on the edge of the enclosure.
NOTE:
Disconnecting the sensor connector plug from the sensor housing will allow for the
complete removal of the internal system from the device enclosure. Disconnecting the
internal system may provide ease in accessing the control board terminals for wiring.
Reconnect the sensor connector plug before reinstalling the internal system.