VOC Pro Transmitter • 55
14. Plug the sensor element into the sensor housing board. Ensure that the pins on the sensing
element align with the sockets on the sensor housing board.
15. Screw the sensor housing cap back onto the sensor housing base, ensuring that the sensor
housing cap is only tightened hand tight.
16. Zero and calibrate the detector as described on page 59.
Sensor Replacement
RKI recommends replacing the sensor element whenever a slow response to gas is observed
during the normal calibration process. After replacing the sensor element, the device MUST
be zeroed and calibrated for proper operation of the device.
WARNING: Do not remove the sensor housing cap or enclosure lid while the circuits are
energized unless the area is determined to be non-hazardous. Keep the sensor
housing cap and enclosure lid tightly closed during operation.
CAUTION:
The internal components can be static sensitive. Use caution when opening the
enclosure and handling internal components. DO NOT use any metal objects or
tools to remove the sensor.
1. Press and hold the SUB button for approximately 6 seconds, until “OFF” shows on the
display screen.
2. Unscrew the rain guard.