36
•
Chapter 6: Gas Type Mode
65-2662XL-04 M2A-XL Transmitter
7. When you have reviewed and accepted the last item in Configuration Mode by pressing the
ENTER button, display will ask
SAVE IT? YES/NO
.
8. To save the configuration settings, press and release the UP/YES button.The display will
indicate
Config Saved
and the M2A-XL will begin its warm-up sequence.
To discard the settings and review them again, press the DOWN/NO button. The display
will ask
DO OVER? YES/NO
. Press the UP/YES button and the display will indicate
Re-
do Config
, then display the target gas and the detection range for a few seconds before
returning to Configuration Mode. Repeat steps 6 - 8 until the settings are as desired.
9. After the M2A-XL completes its warm-up sequence, the display will indicate
CAL
NEEDED
. Since the gas type has been changed, a successful calibration must be performed
before the M2A-XL can enter normal operation.
10. Press and release the UP/YES button to enter Calibration Mode. Normally, you must hold the
UP/YES button for 5 seconds to enter Calibration Mode, but since a calibration is required,
only a momentary push is needed to enter Calibration Mode after changing the gas type.
11. See “Calibration” on page 41 for calibration instructions.
NOTE:
When calibrating an M2A-XL after changing the gas type, the M2A-XL will
not ask if you want to calibrate, or whether you want to perform a fresh air
adjustment, span adjustment or zero adjustment. Since a complete
calibration is required, the calibration sequence will proceed without asking
if you want to perform certain operations.