65-2658XL-CH4-4 M2A-XL Transmitter Operator’s Manual
Chapter 5: Configuration Mode
•
31
5. If you do not wish to save the adjustments and want to exit Configuration Mode, press and
release the DOWN/NO button. The
DO OVER? YES/NO
message will display. Press and
release the DOWN/NO button. The
ABORT? YES/NO
message will display. Press the UP/
YES button to return to normal operation.
If you wish to change some of the adjustments made, press and release the DOWN/NO
button. The
DO OVER? YES/NO
message will display. Press and release the UP/YES
button. The
Re-do Config
message will display and the M2A-XL will return to the first
adjustable parameter. Go back to step 3 and continue.
If you wish to save the adjustments made, press and release the UP/YES button.
Config
Saved
is indicated on the display for a few seconds and the M2A-XL returns to normal
operation.