65-2645-05SS-04 M2A Transmitter Operator’s Manual
31
9. After the M2A completes its warm-up sequence, the display will indicate
CAL NEEDED
. Since the gas
type has been changed, a successful calibration must be performed before the M2A can enter normal
operation.
10. Press and release the UP/YES button to enter Calibration Mode. Normally, you must hold the UP/YES
button for 5 seconds to enter Calibration Mode, but since a calibration is required, only a momentary push is
needed to enter Calibration Mode after changing the gas type.
11. See “Calibration” on page 35 for calibration instructions.
NOTE:
When calibrating an M2A after changing the gas type, the M2A will not ask if you want to calibrate,
or whether you want to perform a fresh air adjustment, span adjustment or zero adjustment. Since a
complete calibration is required, the calibration sequence will proceed without asking if you want to
perform certain operations.