![RKI Instruments 65-2640RK-05-04 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/rki-instruments/65-2640rk-05-04/65-2640rk-05-04_operators-manual_1466975037.webp)
65-2640RK-05-04 M2A Transmitter Operator’s Manual
37
4.
Use the flexible tubing coming from the calibration adapter/splash guard to connect
the regulator to the calibration adapter/splash guard.
5.
Turn the regulator’s on/off knob counterclockwise to open the regulator. Gas will
begin to flow. Take note of the time or start a stopwatch.
6.
When you first start to notice an increase in the gas reading, note how much time
passed between starting the gas flow and noticing a reading response.
7.
Turn the regulator’s on/off knob clockwise to close the regulator.
8.
Unscrew the regulator from the calibration cylinder.
9.
Disconnect the calibration adapter/splash guard’s flexible tubing from the regulator.
NOTE:
Do not disconnect the flexible tubing from the calibration adapter/splash guard.
10. Add 1 minute to the length of time determined in step 6. This is the response time for
the detector. When performing a response test, zero adjustment, or span adjustment,
gas must be applied for this length of time.
Calibration
This section describes how to calibrate the M2A transmitter. It includes procedures to
prepare for calibration, enter Calibration Mode, adjust the fresh air (zero) setting, adjust
the span setting, and return to normal operation.
WARNING:
The M2A is not an active gas monitoring device during the calibration
procedure. The 4-20 mA output signal will “freeze” at 3.5 mA and all
relays will remain in their non-alarm state while the M2A is in
Calibration Mode. The output signal will not indicate current readings
and the relays will not resume operating normally until the M2A is in
normal operation again.
NOTE:
The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a zero air cylinder and a fixed flow regulator with an
on/off knob.
NOTE:
While in the calibration program, if there is no switch activity for the
calibration time-out period the unit will return to normal operation. See
“Viewing & Changing M2A Parameters” on page 28 for instructions to set the
calibration time-out.
NOTE:
The following procedure assumes that the target gas is present in a high
enough concentration to affect the fresh air (zero) reading. If a fresh air
environment can be verified, applying zero air to the detector is not necessary
when setting the zero (span for oxygen) reading.
Non-Intrusive Calibration
If the M2A is installed in a classified area and non-intrusive calibration is required, follow
the instructions below, but do not remove the junction box cover. Use the magnetic wand