20 • 65-2422RK-05 H
2
S Transmitter
NOTE:
For convenience, leave the components of the calibration kit connected by the sample
tubing.
Returning to Normal Operation
1. Remove the voltmeter leads from the amplifier test points.
2. Secure the junction box cover to the junction box.
3. When the display reading falls below the alarm setpoints, return the controller to normal
operation.
4. Verify that the controller display reading decreases and stabilizes at 0 ppm.
5. Store the components of the calibration kit in a safe and convenient place.
Parts List
Table 5 lists replacement parts and accessories for the H
2
S transmitter.
Table 5: Parts List
Part Number
Description
06-1248RK
Sample tubing (order by the foot)
07-0033RK
Detector housing cap gasket
07-0203RK
Rubber retaining boot
14-2101RK
Spacer between sensor and rubber boot
18-0405RK-01
Junction box (without cover; pre drilled for amplifier)
18-0406RK
Junction box cover
57-1060RK
Amplifier (specify target gas when ordering)
65-2422RK-05
H
2
S transmitter (includes detector and amplifier), CSA classified
65-2428
H
2
S replacement detector assembly (includes sensor), CSA classified
71-0113RK
65-2422RK-05 H
2
S Transmitter Operator’s Manual
(this document)
81-0076RK-01
Zero air calibration cylinder, 34 liter steel
81-0151RK-04
Calibration cylinder, 25 PPM H
2
S in nitrogen, 34 liter aluminum
81-1050RK
Regulator with gauge and knob, 0.5 LPM, for 17 liter and 34 liter steel calibration cyl
-
inders (cylinders with external thread)
81-1051RK
Regulator with gauge and knob, 0.5 LPM, for 34 liter aluminum calibration cylinders
and 58 liter and 103 liter steel calibration cylinders (cylinders with internal thread)
81-1117RK
Calibration cup
ES-1537-H2S
H
2
S replacement sensor