background image

 

65-2396RKSS CO

2

 Transmitter • 7

To distinguish the different ranges of CO

2

 detectors from one another, a short length of shrink tubing is 

applied to the wiring where it comes out of the nipple. The following table indicates the color of the 
shrink tubing and the color of the wire to which it is applied.

Amplifier

The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that corresponds 
to the detection range and transmits the signal to a gas monitoring controller. A foam gasket that 
orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. The 
amplifier includes the detector terminal strip, controller terminal strip, span pot, zero pot, and test 
points (see Figure 1).

Controller Terminal Strip

The controller terminal strip is a three position plug-in style terminal strip located near the top of the 
amplifier. Use the controller terminal strip to wire the CO

2

 transmitter to a controller.

Detector Terminal Strip

The detector terminal strip is a four position plug-in style terminal strip located below the controller 
terminal strip. Use the detector terminal strip to connect the CO

2

 detector to the amplifier.

NOTE:

The IR CO

2

 detector is factory-wired to the detector terminal strip. See “Wiring the CO

2

 

Transmitter to a Controller” on page 9 for all wiring procedures related to the transmitter.

Zero Pot

The zero pot is located in the upper left corner of the amplifier (see Figure 1). Use a small flat blade 
screwdriver to turn the zero pot’s adjustment screw and adjust the amplifier’s target gas-free output 
during the start-up and calibration procedures. Turn the adjustment screw clockwise to increase the 
zero output and counterclockwise to decrease the zero output.

Span Pot

The span pot is located to the right of the zero pot (see Figure 1). Use a small flat blade screwdriver to 
turn the span pot’s adjustment screw and adjust the amplifier’s gas response output during the 
calibration procedure. Turn the adjustment screw clockwise to increase the gas response and 
counterclockwise to decrease the gas response.

Table 2:CO

2

 Detector Color Designations

Detector

Color Designations

0 - 5,000 ppm

black shrink tubing on white wire

0 - 9,000 ppm

white shrink tubing on red wire

0 - 5% volume

green shrink tubing on white wire

0 - 50% volume

green shrink tubing on red wire

0 - 100% volume

red shrink tubing on green wire

Содержание 65-2396RKSS Series

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2396RKSS CO2 Transmitter Operator s Manual Part Number 71 0338 Revision 0 Released 9 1 20...

Страница 2: ...calibration and maintenance of the transmitter is essential for proper operation and correct readings Please calibrate and maintain this transmitter regularly Frequency of calibration depends upon the...

Страница 3: ...ent costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPR...

Страница 4: ...iring the CO2 Transmitter 9 Startup 12 Introducing Incoming Power 12 Setting the Zero Signal 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 14 Replacing Components of the CO2 Transmitter...

Страница 5: ...in this manual to assure proper and safe operation of the 65 2396RKSS and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 65 2396RKSS as described in this m...

Страница 6: ...a 1 1 4 hex that allows removal or installation of the detector with a wrench A porous flame arrestor that is coated with a hydrophobic film that repels liquids is on one end of the detector and allow...

Страница 7: ...minal strip Use the detector terminal strip to connect the CO2 detector to the amplifier NOTE The IR CO2 detector is factory wired to the detector terminal strip See Wiring the CO2 Transmitter to a Co...

Страница 8: ...connect wiring from the amplifier to the controller An O ring seals the interface between the junction box cover and the junction box base The detector and amplifier are factory installed in the junct...

Страница 9: ...ing down see Figure 2 Wiring the CO2 Transmitter to a Controller WARNING Always verify that the power to the controller is off before you make wiring connections 1 Turn off the controller 2 Turn off o...

Страница 10: ...re 3 Connect the positive wire to the PWR SIG terminal Connect the signal wire to the PWR SIG S terminal Connect the negative wire to the PWR SIG terminal CAUTION If shielded cable is used leave the c...

Страница 11: ...an available chassis earth ground at the controller RKI controllers typically have a ground stud that can be used to ground the cable s drain wire 24 VDC DC Ground 4 20 mA In S Amplifier Controller C...

Страница 12: ...ince there is a background of CO2 in air of typically 300 600 ppm 0 03 0 06 volume it is necessary to use a calibration kit with a 100 nitrogen cylinder to set the zero signal of a CO2 transmitter wit...

Страница 13: ...junction box Maintenance This section describes maintenance procedures It includes preventive maintenance troubleshooting and component replacement procedures Preventive Maintenance This section descr...

Страница 14: ...ring is correct and secure 2 Calibrate the transmitter 3 If the fail condition continues replace the detector 4 If the fail condition continues contact RKI for further instruction Slow or No Response...

Страница 15: ...about a one minute period when the 4 20 mA output is stabilizing and may be above the controller alarm points or well below zero momentarily RKI controllers have a one minute warmup period when the c...

Страница 16: ...of the zero pot Be sure to make arrangements so that this does not cause unwanted alarms 10 Turn on power to the controller 11 Turn on the controller and place it into normal operation NOTE When first...

Страница 17: ...n set the zero reading set the response reading and return to normal operation WARNING The controller is not an active gas monitoring device during the calibration procedure NOTE The following procedu...

Страница 18: ...ator into the calibration cylinder Verify that the calibration gas is representative of the transmitter s target gas 2 Turn the regulator s on off knob counterclockwise to open the regulator 3 Allow t...

Страница 19: ...tor 0 100 Volume 65 2396RKSS 01 Infrared CO2 transmitter 0 9 000 ppm includes detector junction box and amplifier 65 2396RKSS 02 Infrared CO2 transmitter 0 5 000 ppm includes detector junction box and...

Страница 20: ...with gauge and knob 0 5 liter minute continuous flow for 17 liter and 34 liter steel calibration cylinders cylinders with external threads 81 1051RK Regulator with gauge and knob 0 5 liter minute con...

Отзывы: