65-2394RK-CH4-4 Combustible Gas Transmitter • 17
Preparing for Calibration
1.
Unscrew and remove the amplifier junction box’s cover from the junction box.
2.
Set a voltmeter to measure in the millivolt (mV) range.
3.
Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead
into the red (+) test point; plug the negative lead into the black (-) test point.
4.
Use the following formula to determine the correct test points output for the
calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50 %LEL and a fullscale setting of
100 %LEL, the correct output is 300 mV.
300(mV) = (50/100) X 400 +100
5.
Place the controller into its calibration mode or disable external alarms.
NOTE:
Calibrating the combustible gas transmitter may cause alarms. Be sure to put the
controller into its calibration mode or disable external alarms before continuing.
Setting the Zero Reading
NOTE:
If you can verify that the combustible gas transmitter is in a fresh air
environment, you do not need to apply zero air to the detector before adjusting
the zero reading.
1.
Screw the regulator into the zero air calibration cylinder.
2.
Use the flexible tubing coming from the calibration cup/splash guard to connect the
regulator to the calibration cup.
3.
Turn the regulator’s on/off knob counterclockwise to open the regulator.
4.
Allow the gas to flow for the length of time determined in “Determining Response
Time” on page 16 and verify a reading of 100 mV (±2 mV). If necessary, use the zero
pot on the amplifier to adjust the reading to 100 mV (±2 mV).
5.
Turn the regulator’s on/off knob clockwise to close the regulator.
6.
Unscrew the regulator from the zero air calibration cylinder.
7.
Leave the calibration cup/splash guard’s flexible tubing connected to the regulator.
Setting the Response Reading
1.
Screw the regulator into the calibration cylinder. Verify that the calibration gas is
representative of the transmitter’s target gas.
2.
Turn the regulator’s on/off knob counterclockwise to open the regulator.
3.
Allow the calibration gas to flow for the length of time determined in “Determining
Response Time” on page 16 and verify that the reading matches the response reading
(±2 mV) you determined earlier. If necessary, use the span pot on the amplifier to
adjust the reading to match the correct response reading.
4.
Turn the regulator’s on/off knob clockwise to close the regulator.
5.
Unscrew the regulator from the calibration cylinder.
www.
GlobalTestSupply
.com
Find Quality Products Online at: