35-3010RK-05 Sample-Draw Detector • 1
Overview
This manual describes the 35-3010RK-05 sample draw detector. This manual also
describes how to install, start up, maintain, and calibrate the detector head when it is used
with a gas monitoring controller. A parts list at the end of this manual lists replacement
parts and accessories for the detector head.
Specifications
Table 1 lists specifications for the 35-3010RK-05.
WARNING: When using the 35-3010RK-05, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the
35-3010RK-05 and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the 35-3010RK-05 as described in this
manual.
Table 1: Specifications
Target Gases & Detection Ranges
Oxygen: 0 - 25%
Carbon Monoxide (CO): 0 - 300 ppm
Input Power
24 VDC
Current Draw
200 mA
Construction (housing)
Fiberglass/polyester (NEMA 4X)
Dimensions
15.44 in. H x 12.55 in. W x 8.31 in. D
Weight
14 lbs.
Sampling Method
Sample draw
Sample Flow
1.2 SCFH (nominal)
Response Time
90% in 30 seconds
Accuracy
Oxygen:
± 0.5% O
2
Carbon Monoxide:
± 5% of reading or ± 5 ppm CO (whichever is greater)