10 • 35-3001A-14 Methane Sample-Draw Detector
8. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in
Figure 5. Refer to the controller operator’s manual and the controller detector head specification
sheet for the 35-3001A-14 for detector/terminal strip connections specific to the controller.
Figure 5: Wiring the Sample-Draw Detector to a Controller
9. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the
controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain
wire.
Whi te
Not Used on
This Version
Black
Red
Z ER O
S PA N
S
P W R /S IG
Whi te
Red
Green
Black
Red
LEL/I R
LEL/IR
AMP 2
AMP 1
AMP 1
AMP 2
PATS201PATS201
Whi te
PATS101PATS101
OXY
IR CH4
Sensor PCB
IR CH4
Sensor
OXY
S ensor
C urr ent
148 mA
IR CH4
Sensor PCB
S PA N
Black
W
L E L
G
B
R
Z ER O
S
P W R / SIG
W
L EL
G
B
R
S ensor
C urr ent
148 mA
Red
Black
Green
4/20 Signal
+ 24 VDC
Whi te
4/20 Signal
+ 24 VDC
IR CH4
Sensor
Controller or Recording Device
- (DC Ground)
- (DC Ground)
CH4 2
CH4 1