background image

 

12 • 35-3001A-01H Hydrogen Sample-Draw Detector

6. If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until the 

voltmeter reading is 100 mV (± 2 mV).

7. Remove the voltmeter leads from the test points.
8. Close the housing door.

Maintenance

This section describes maintenance procedures. It includes preventive maintenance procedures. 
This section also includes procedures to troubleshoot the sample-draw detector, replace 
components of the sample-draw detector, adjust the low flow setting, and remove the particle 
filter’s tubing stub.

Preventive Maintenance

This section describes a preventive maintenance schedule to ensure the optimum performance of 
the sample-draw detector. It includes daily, monthly, and quarterly procedures.

Daily Visual Checks

1. Verify that the Pilot LED is on.

2. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary use the 

sensor flow control valve to adjust the flow rate to 1.0 SCFH. Turn the valve’s knob 
counterclockwise to increase the flow and clockwise to decrease the flow.

3. Verify a display reading of 0 %LEL at the controller or a reading of 100 mV at the amplifier 

test points. Investigate significant changes in the display reading.

Monthly Response Test

This procedure describes a test to verify that the sample-draw detector responds properly to the 
target gas.

NOTE:

To reduce the response time of this test, use a short incoming sample line. If the sample-
draw detector’s sample line is long, connect a shorter line for this test. Make sure you 
reconnect the sample line after you complete this procedure.

NOTE:

Performing a response test on the sample-draw detector may cause alarms. Be sure to put 
the controller into its calibration program or disable external alarms before performing 
this test.

Preparing for the response test

NOTE:

This procedure describes the RKI calibration kit that includes a demand flow regulator.

1. Verify that the display reading at the controller is 0 % LEL.

If the display reading is not 0 % LEL, set the zero reading as described in “Calibration” on 
page 18, t
hen continue this procedure.

2. Set a voltmeter to measure in the millivolt (mV) range.

3. Open the housing door, then plug the voltmeter leads into the test points on the amplifier.

Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test 
point.

Содержание 35-3001A-01H

Страница 1: ...www rkiinstruments com 35 3001A 01H Hydrogen Sample Draw Detector Operator s Manual Part Number 71 0307RK Revision A Released 8 7 17 ...

Страница 2: ...eriodic calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type of use you have and the sensor types Typical calibration frequencies for most applications are between 3 and 6 months but can be required more often or less often based on your usage ...

Страница 3: ...eplacement costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRESSED OR IMPLIED AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS INC INCLUDING BUT NOT LIMITED TO THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IN ...

Страница 4: ... Draw Detector to a Controller 9 Start Up 11 Introducing Incoming Power 11 Setting the Zero Reading 11 Maintenance 12 Preventive Maintenance 12 Troubleshooting 13 Replacing Components of the Hydrogen Sample Draw Detector 14 Adjusting the Low Flow Setting 16 Removing the Particle Filter s Tubing Stub if Necessary 17 Calibration Frequency 18 Calibration 18 Preparing for Calibration 18 Setting the Ze...

Страница 5: ...r and safe operation of the 35 3001A 01H and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 35 3001A 01H as described in this manual Table 1 Specifications Target Gas and Detection Range Hydrogen 0 100 LEL Input Power 24 VDC Nominal 18 5 VDC 30 VDC Construction housing Fiberglass polyester NEMA 4X Dimensions 8 5 in H x 6 5 in W x 4 25 in D Weight 4 5 lbs...

Страница 6: ...ot LED Fail LED Sensor Flow Control Valve Interconnect Terminal Strip Pump Reset Switch Flowmeter W S AMP 1 AMP 2 OXY S R W G G Hydrophobic Filter Particle Filter B LEL IR Flowmeter Circuit Board Connector Inlet Fitting Exhaust Fitting This End To Inlet Fitting 3 4 Conduit Hub Hydrogen Sensor INLET EXHAUST Pump Connector Hydrogen Amplifier S ens or Current 115m A W L EL G B R AMP 2 AMP W B Pressur...

Страница 7: ...ple fittings accept 1 4 in rigid tubing See the Installation section on page 7 to connect tubing to the sample fittings Particle Filter A particle filter with a tubing stub on one end is shipped with the instrument but it is not factory installed If the particle filter is installed directly to the inlet fitting the tubing stub must be used If the particle filter is installed somewhere else like at...

Страница 8: ...ht corner of the main circuit board Use the pump connector to connect the pump to the main circuit board NOTE The pump is factory wired to the main circuit board See Installation on page 7 for all wiring procedures related to the sample draw detector Flowmeter Circuit Board Connector The flowmeter circuit board connector is a six position connector in the upper left corner of the main circuit boar...

Страница 9: ...er The hydrophobic filter is located toward the bottom left of the main circuit board The filter prevents particulates and water in the incoming gas sample from damaging the flow and detection systems Replace the filter when it appears dirty discolored or clogged Pump The pump is mounted to the right side of the main circuit board The pump pulls the gas sample into the sample draw detector The pum...

Страница 10: ...an pot is located below the zero pot see Figure 3 Use a small flat blade screwdriver to turn the span pot s adjustment screw and adjust the amplifier s gas response output during the calibration procedure Turn the adjustment screw clockwise to increase the span output and counterclockwise to decrease the span output CAUTION The amplifier includes additional pots They are factory set Do not adjust ...

Страница 11: ...owing when you select the mounting site Is there enough room to open the housing door and make wiring and tubing connections at the bottom of the housing Make sure there is sufficient room to perform start up maintenance and calibration procedures Are the flowmeter and status LEDs visible Figure 4 Outline and Mounting Dimensions Ø 30 x 50 4X MOUNTING 78 1 13 40 Pilot Fail 85 80 4 50 NOTE Housing i...

Страница 12: ...ng components 4 Connect a length of sample tubing to the other side of the particle filter and route it to the sampling area 1 4 O D rigid polypropylene Teflon or flexible polyurethane tubing may be used RKI Instruments Inc recommends using either 1 4 O D x 1 8 I D or 1 4 O D x 0 170 I D tubing based on your length requirements See Specifications on page 1 for maximum tubing lengths based on tubin...

Страница 13: ...t to seal the connection between the tubing and the inlet fitting 3 Hand tighten the nut on the exhaust fitting so the ferrules clamp on the tubing If the tube nut is removed see Figure 6 for the arrangement of the fitting components 4 Route the opposite end of the tubing to an open area where the sample can safely disperse or to an exhaust duct See Specifications on page 1 for maximum tubing leng...

Страница 14: ... 5 Wiring the Sample Draw Detector to a Controller 9 If shielded cable is used connect the cable s drain wire to an available chassis earth ground at the controller RKI controllers typically have a ground stud that can be used to ground the cable s drain wire Not Used on This Version S P WR SI G ZE RO SP AN OXY LEL IR Red White AMP 1 P A T S 2 0 1 P A T S 2 0 1 LEL IR 24 VDC AMP 1 OXY Controller o...

Страница 15: ...ing this test 7 Verify that the incoming sample line is not leaking To test the sample line press and hold the reset switch and plug the open end of the sample line with your thumb If the flowmeter ball drops to the bottom of the flowmeter the incoming sample line is not leaking 8 Remove your thumb from the sample line release the reset switch and verify the flowmeter returns to a normal flow rate...

Страница 16: ...display reading of 0 LEL at the controller or a reading of 100 mV at the amplifier test points Investigate significant changes in the display reading Monthly Response Test This procedure describes a test to verify that the sample draw detector responds properly to the target gas NOTE To reduce the response time of this test use a short incoming sample line If the sample draw detector s sample line...

Страница 17: ... safe place Quarterly Calibration Calibrate the sample draw detector as described in Calibration on page 18 Troubleshooting The troubleshooting guide describes symptoms probable causes and recommended action for problems you may encounter with the sample draw detector NOTE This troubleshooting guide describes sample draw detector problems only See the controller Operator s Manual if the controller...

Страница 18: ...sed use a different demand flow regulator to determine if the original one is functioning properly 3 If necessary set the correct flow rate with the sensor flow control valve 4 If you cannot set the correct flow rate check the sample line for obstructions or kinks 5 If the calibration response difficulties continue replace the sensor as described later in this section 6 If the calibration response...

Страница 19: ...ier and remove the screw lock washer and flat washer Be careful not to lose any of these parts 6 Remove the old amplifier from the main circuit board 7 Install the new amplifier in the same orientation as the old amplifier See Figure 1 8 Reinstall the screw lock washer and flat washer you removed in step 5 9 Install the detector and controller plug in terminals strips into their sockets on the new...

Страница 20: ... that the flow rate is approximately 1 0 SCFH then close the housing door Replacing the Particle Filter 1 Turn off the controller 2 Turn off power to the controller 3 If the particle filter is installed at the inlet fitting 1 Disconnect the tubing routed to the sampling area if installed from the particle filter 2 Disconnect the particle filter from the tubing stub 3 Install the new particle filte...

Страница 21: ... until the unit is out of low flow alarm when the reset switch is pressed and released 4 Decrease the flow very slowly by turning the sensor flow control valve s knob clockwise and verify that the low flow alarm is 0 6 SCFH 0 1 If the low flow alarm is set too high turn the pressure switch adjustment screw slightly clockwise Repeat steps 3 and 4 if necessary 5 Use the sensor flow control valve to ...

Страница 22: ...y 3 to 6 months is adequate for most applications Unless experience in a particular application dictates otherwise RKI Instruments Inc recommends a calibration frequency of every 3 months If an application is not very demanding for example detection in a clean temperature controlled environment where combustible gases are not normally present and calibration adjustments are minimal at calibration ...

Страница 23: ...ep 5 If you verified a fresh air environment proceed to the next section Setting the Response Reading 5 Disconnect the sample tubing from the inlet line 6 Unscrew the regulator from the zero air calibration cylinder Setting the Response Reading 1 Screw the regulator into the hydrogen calibration cylinder 2 Connect the sample tubing from the demand flow regulator to the sample draw detector s inlet...

Страница 24: ...0167RK Particle filter 57 1053RK 01 Amplifier specify target gas when ordering 61 0155RK Hydrogen sensor plug in type LEL range 71 0307RK Operator s Manual 35 3001A 01H Sample Draw Detector 81 0002RK 01 Calibration cylinder 50 LEL hydrogen in air 34 liter steel 81 0002RK 03 Calibration cylinder 50 LEL hydrogen in air 103 liter 81 0076RK 01 Zero air calibration cylinder 34 liter steel 81 0076RK 03 ...

Отзывы: