Chapter 5: User Mode and Calibration • 49
14. See “Exiting the GAS CAL Menu” on page 49 to return to User Mode.
Exiting the GAS CAL Menu
1. While in the
GAS CAL
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
GAS CAL
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 62 to enter Measuring Mode.
Setting Calibration Parameters (CAL SET)
The
CAL SET
menu has 4 items:
CAL.RMDR
,
CAL.INT
,
CAL.EXPD
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
CAL SET
.
2. Press and release POWER MODE. The
CAL.RMDR
item appears.
CAL.RMDR
ON
(factory setting): The SC-04 gives an indication at start up if it is due for calibration. The
type of indication depends on the
CAL.EXPD
setting (see pg.50).
OFF
: The SC-04 does not give an indication at start up if it is due for calibration.
1. After entering the
CAL SET
menu, press AIR to scroll to
CAL.RMDR
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
CAL.RMDR
item
in the
CAL SET
menu.
5. See “Exiting the CAL SET Menu” on page 50 to return to User Mode.
CAL SET
CAL .RMDR
CAL .RMDR