![RJG Sensor PreCheck Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/rjg/sensor-precheck/sensor-precheck_user-manual_1466499009.webp)
RJG, INC. SENSOR PRECHECK | QUICK START GUIDE
3
RUN AN AUTOMATED SENSOR TEST
Select
a
1
Sensor PreCheck
icon from the application
home page with which to connect. The Sensor PreCheck
icon will be grey until selected, and will turn green after
selection. If multiple sensors are attached for testing,
a pop‑up window will appear to determine if each
connected sensor is the same or different model(s).
Select
the appropriate response for the connected
sensor(s) to continue.
Enter the
2
Mold Number
in the provided field.
Select
3
Begin Automated Test
to test the sensor(s). The
4
Progress Bar
will indicate the test progress. Wait for the
test to complete.
The
5
automated test complete page
will indicate any
alerts, warnings, and the number of sensors that passed
the test.
Select
a
6
sensor
to enter the sensor
information, including Location, Cavity Name, Pin Size,
Sensitivity, Sensor Model, and Sensor Serial Number.
Complete the testing by running a force test (page
4).