![Rishabh Instruments RISH Master 3430i Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/rishabh-instruments/rish-master-3430i/rish-master-3430i_operating-manual_1464364044.webp)
is from 2 to 85.
Default value is 42.
This address allows to read or set the value of
contrast of display LCD. The valid range is from 6 to 23.
Default value is 9.
Default value for phase 1 : 160, 82, 45; phase 2 : 0, 0, 0
and phase 3 : 128, 128, 128.
Address
Parameter
Description
R
40041
40081
Word Order
Energy update
rate
Energy update rate is the time after which Energy registers are
updated. This time is user settable from 1 - 60 minutes
Word Order controls the order in which the
receives
instrument
or sends floating - point numbers:- normal or reversed register
order.In normal mode, the two registers that make up a floating
point numbers are sent most signicant bytes rst.
In reversed register mode , the two registers that make up
a floating point numbers are sent least signicant bytes rst.
To set the mode, write the value ‘2141.0’ into this register-
the instrument will detect the order used to send this value and
set that order for all ModBus transaction involving floating point
numbers.
40085
40089
to
40105
RGB Color
Code for
L1, L2, L3
This addresses allow to read or set the value of Red,
Green, Blue component of color used to display
phase 1, phase 2, phase 3 parameters respectively.
The valid range is 0 to 255.
40087
42