otherwise it will turn green. The go no go settings can be changed by pressing the
icon on the settings window.
You can change the go no go insulation resistance limit by clicking on the go no go
insulation resistance button after which a keyboard will appear where you can insert the
desired value and press the button. The unit of the insulation resistance can be changed by
pressing the go no go resistance unit button. Similarly you can also change the go no go
voltage limit by pressing on the go no go voltage button and entering the desired value on the
keyboard and then pressing the button. You can also turn on/off go no go by clicking on/off
button shown in the above window. The icon of this button shows whether the go no go
function is on/off.
9.7 Other Settings:
Apart from all the above settings there are also a few more additional settings. These
include the Communication interface setting, Audio on/off feature, and the reset default
settings button which will reset control settings, and the other settings in the settings menu
to their default value. To change these settings in the settings window, press
the icon. Then the more settings window will be displayed as shown on the next page.
ir limit
button
ir limit unit
button
voltage
limit button
43