M-UM008-01 Page 3

Caring for the DLS 350 Output Port

Using the 2kHz Function

The DLS 350 utilizes a physical fiber connection at the output port. This ensures a steady 

power level for performing insertion loss tests.

Be aware that any test jumpers must be inspected and cleaned before plugging into the 

DLS 350 unit. If soiled or damaged connectors are inserted, they can cause damage to the 

DLS 350 output port and the unit may need to be repaired. 

When the 2kHz modulation is active on the DLS 350, the currently-selected wavelength 

indicator will blink. The 2kHz modulation will be recognized by optical power meters further 

down the fiber and indicated by a 2kHz notification onscreen and a loud beep.

The test jumper has 

physical contact with the 

ferrule inside the DLS 350.

To clean the ferrule  

inside the DLS 350:

• 

Unplug the test jumper

• 

Unscrew the adapter until 

it spins freely, then pull

• 

Inspect the ferrule with 

an approved microscope, 

clean if necessary, and 

replace adapter