TECHNICAL DATA 

(

According to ANSI standard S3.22 2003

)

Maximum-OSPL90

125 dB (1000 Hz)

HFA-OSPL90

122 dB

HFA-full-on Acoustic Gain

55 dB

Reference Test Gain

45 dB

Frequency Range

200 Hz to 5400 Hz

Equivalent Input Noise Level

21 dB

Total Harmonic Distortion

500 Hz: 2%

800 Hz: 2% 1600 Hz: 1%

AGC

CRC (Input AGC)/OPC (Output AGC)

Operating Switch

O-M2-M

Output Limiting Control

OPC (range: 18 dB)

Tone Control

TONE H, TONE L

Gain Control

Main VR (range: 30 dB)

GAIN (range: 34 dB)

Battery Type/Supply Voltage

13/1.3 V

Battery Current

0.94 mA

Battery Life

Approx. 280 hours

Dimensions/Weight

3.75 × 1.27 × 0.88 cm/4.0 g

 

(excluding battery)

(Typical value with E1 hook)

TECHNICAL DATA

 

(

)

Reference Test Frequency 1600 Hz

OSPL

90

125 dB

500 Hz

122 dB

Peak

132 dB

Full-on Acoustic Gain

58 dB

Equivalent Input Noise Level

22 dB

Total Harmonic Distortion

500 Hz: 3% 800 Hz: 2% 1600 Hz: 2%

AGC

CRC (Input AGC)/OPC (Output AGC)

Operating Switch

O-M2-M

Output Limiting Control

OPC (range: 18 dB)

Tone Control

TONE H, TONE L

Gain Control

Main VR (range: 30 dB)

GAIN (range: 34 dB)

Battery Type/Supply Voltage

PR48/1.3 V

Battery Current

1.0 mA

Battery Life

Approx. 260 hours

Dimensions/Weight

3.75 × 1.27 × 0.88 cm/4.0 g

 

(excluding battery)

(Typical value with E1 hook)

z

Memory selection (using selector switch)
I f   s o u n d s   a r e   h a r d   t o   h e a r
depending on the environment,
s u c h   a s   i n   a   n o i s y   o r   q u i e t
environment, change the setting of
the memory selector to find the
position that yields best results.

* The memory selector switch (M2/

M) is programmed to match your
hearing pattern. Operate the switch
to find the position that yields best
results in a given situation.

z

Switch operation sound
Memory 1

Memory 2

M2

Æ

M: one beep

Æ

 M2: two beeps

Removing Hearing Aid

1. Set the switch to "O" (power off) before

removing the hearing aid. If the switch is left

No. 40240 04-10

in the "M" or "M2" position, acoustic feedback (howling) may
occur.

2. Be sure to set the switch to "O"

 

and take out the battery,

when not using the hearing aid.

Turn-on precaution

If you switch the hearing aid off and then immediately on again,
there may be no sound, but this is not a defect. In such a case,
return the switch to "O" (Off), wait for at least 10 seconds, and
then set the switch to "M" or "M2" again.

Adjusting the Controls

This hearing aid is a digital type. To adjust it, a computer (with
NOAH and the Rionet Selector installed) and a HI-PRO unit are
required. Have the adjustment done at the dealer.

Factory Default Setting

z

O - M2 - M switch :

"O"

z

Volume control (1 - 4) : "1"

According to IEC standard Pub. 60118-0-1983
Amendment 1-1994

Full-on gain curve

Frequency response curve and
effect of tone control

OSPL90 curve and effect of
output limiting control

Full-on acoustic gain
frequency response curve

Basic frequency response curve
and effect of tone control

OSPL

90

 curve and effect of

output limiting control

Отзывы: