rionet HB-G4P Скачать руководство пользователя страница 2

  O-

M

T-M Switch

O:  Hearing aid is turned off and no sound is heard.
MT:  Sounds through both the microphone and telephone coil can 

be heard. Set to "MT" position when using your hearing aid 
for telephone or induction loop system in combination with 
the MT balancer.

When the switch of the hearing aid is set to MT, noise may be 
heard in the vicinity of electromagnetic fields, such as generated 
for example by electronically secured gates or similar.

M:  Microphone for ordinary use. This is the normal operating 

position.

Removing Hearing Aid

1.  Set the switch to "O" (power off) before re-

moving the hearing aid. If the switch is left in 
the "M" or "MT" position, acoustic feedback 
(high-pitched sounds) may occur.

2. Be sure to set the switch to "O"

 

and take out the battery, when 

not using the hearing aid.

Recycled papers are used for this manual.

No. 51310  06-12

Turn-on precaution

If you switch the hearing aid off and then immediately on again, 
there may be no sound, but this is not a defect. In such a case, 
return the switch to "O" (Off), wait for at least 10 seconds, and 
then set the switch to "M" or "MT" again.

Adjusting the Controls 

This hearing aid is a digital type. Any adjustments must be done 
at the dealer. For details, please contact the dealer.

Factory default setting for communication/Audio input selector

The default setting is as shown at right. 

FIT:  Position for hearing aid fitting adjustment 
 

:  Position for Audio input 

Other setting

  O - MT - M switch: 

"O"

  Volume control (1 to 4):  "1"

140

0dB=20µPa

1600 2500

130

120

110

100

90

100

200

500

1000

2000

5000

10000

Frequency (Hz)

O

ut

pu

tS

P

L

(d

B

)

Input SPL 60dB

150

0dB=20µPa

1600 2500

140

130

120

110

100

100

200

500

1000

2000

5000

10000

Frequency (Hz)

O

ut

pu

tS

P

L

(d

B

)

Input SPL 90dB

0 dB

-19 dB

80

1600 2500

70

60

50

40

30

100

200

500

1000

2000

5000

10000

Input SPL 50dB

(0dB=20µPa)

Frequency (Hz)

A

co

us

tic

G

ai

n

(d

B

)

140

0dB=20µPa

1600 2500

130

120

110

100

90

100

200

500

1000

2000

5000

10000

Frequency (Hz)

O

ut

pu

tS

P

L

(d

B

)

Input SPL 60dB

150

0dB=20µPa

1600 2500

140

130

120

110

100

100

200

500

1000

2000

5000

10000

Frequency (Hz)

O

ut

pu

tS

P

L

(d

B

)

Input SPL 90dB

0 dB

-19 dB

80

1600 2500

70

60

50

40

30

100

200

500

1000

2000

5000

10000

Input SPL 50dB

(0dB=20µPa)

Frequency (Hz)

A

co

us

tic

G

ai

n

(d

B

)

(off) (on) (on)

O M T M

Full-on gain curve

Frequency response curve 

OSPL90 curve and effect of 
output limiting control

Full-on acoustic gain 
frequency response curve

Basic  frequency  response 
curve

OSPL

90

  curve  and  effect 

of output limiting control

Note:

(1)  Production  number  (se-

rial number) is indicated at 

the upper rear side of the 

unit.

(2) Manufacturer's  name  is 

imprinted at the lower rear 

side of the unit.

TECHNICAL DATA 

(

According to ANSI standard S3.22 2003

)

Maximum-OSPL90

141 dB (850 Hz)

HFA-OSPL90

133 dB

HFA-full-on Acoustic Gain   66 dB
Reference Test Gain

  56 dB

Frequency Range

200 Hz to 5000 Hz

Equivalent Input Noise Level

 

25 dB

Total Harmonic Distortion  500 Hz: 4%  800 Hz:1%  1600 Hz: 2%
HFA-SPLIV

113 dB at 31.6 mA/m

AGC

Input AGC / OPC (Output AGC)

Attack / Recovery Time

Input AGC

Attack / Recovery: 4 or 252 msec 

Output AGC

Attack: 0.25 msec

Recovery: 0.25 or 32 msec

Operating Switch

O-MT-M

Output Limiting Control

OPC (range: 19 dB)

Gain Control

Main VR (range: 16 dB)
GAIN (range: 19 dB)

Battery Type/Supply Voltage 675/1.3 V
Battery Current

2.7 mA

Battery Life

Approx. 160 hours

Dimensions/Weight

4.67 × 1.62 × 0.97 cm /

6.7 g (excluding battery)

(Typical value with E0 hook)

TECHNICAL DATA 

(

 

)

Reference Test Frequency 1600 Hz
OSPL

90

138 dB

500 Hz

137 dB

Peak

145 dB

Full-on Acoustic Gain

  75 dB

Equivalent Input Noise Level   24 dB
Total Harmonic Distortion 500 Hz: 2%  800 Hz: 1%  1600 Hz: 2%
Induction Coil Sensitivity

105 dB at 1 mA/m

AGC

Input AGC / OPC (Output AGC)

Attack / Recovery Time

Input AGC

Attack / Recovery: 4 or 252 msec

Output AGC

Attack: 0.25 msec

Recovery: 0.25 or 32 msec

Operating Switch

O-MT-M

Output Limiting Control

OPC (range: 19 dB)

Gain Control

Main VR (range: 16 dB)
GAIN (range: 19 dB)

Battery Type/Supply Voltage PR44/1.3 V
Battery Current

2.7 mA

Battery Life

Approx. 160 hours

Dimensions/Weight

4.67 × 1.62 × 0.97 cm /

6.7 g (excluding battery)

(Typical value with E0 hook)

According to IEC standard Pub. 60118-0-1983 

Amendment 1-1994