![rionet HA-72P Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/rionet/ha-72p/ha-72p_instruction-manual_1463526002.webp)
- S - Switch
: Power off
Set to " " when the hearing aid is not in use.
S
: Power on and noise suppressor will be acti-
vated according to noise levels.
Set to "S" when using the hearing aid in
noisy place.
: Microphone
Set to " " for ordinary use.
When taking off the hearing aid, set the switch to " ". If not,
acoustic feedback may occur.
M-MT Switch
M: Microphone
Set to "M" to hear sounds through the microphone.
MT: Microphone and Telephone Coil
Sounds through both the microphone and telephone
coil can be heard. Set to "MT" position when using
your hearing aid for telephone or induction loop
system in combination with the MT Balancer.
When the switch of the hearing aid is set to MT, noise may be
heard in the vicinity of electromagnetic fields, such as generated
for example by electronically secured gates or similar.
No. 41591 08-01
Volume Control
Adjust the sound volume to your comfortable level with the vol-
ume control.
The best position of the control will depend upon the distance
and voice of the person to whom you are talking; readjustment
may be necessary.
When using the hearing aid for the first time, set the control for
a minimum sound level to hear, then increase the volume gradu-
ally.
1. Set to "4" center click for ordinary use.
Frequency response curve
and effect of tone control
SSPL90 curve and effect of
output limiting control
Full-on gain curve
Basic frequency response curve
and effect of tone control
OSPL
90
curve and effect of
output limiting control
Full-on acoustic gain
frequency response curve
Note:
(1) Production number (se-
rial number) is printed
in the battery compart-
ment.
(2) Manufacturer's name
is shown inside of a
hearing aid.
2008
EC REP
RION CO., LTD.
3-20-41 Higashimotomachi Kokubunji,
Tokyo 185-8533 Japan.
RION CO., LTD. Europe Representative Office
Schaepmanlaan 66 NL-4623 XZ Bergen op Zoom
THE NETHERLANDS
TECHNICAL DATA
(
According to ANSI Standard S3.22 1987
)
Maximum SSPL90
133 dB
HFA-SSPL90
131 dB
HFA-full-on Gain
69 dB
Reference Test Gain
114 dB
Frequency Range
100 Hz to 5500 Hz
Equivalent Input Noise Level
19 dB
Total Harmonic
Distortion
500 Hz
9%
800 Hz
12%
1600 Hz
4%
Induction Coil Sensitivity
126 dB at 10 mA / m
Operating Switch
- S - / M-MT
Output Limiting Control
MOP, range 20 dB (continuous)
Tone Control
TONE H, TONE L (continuous)
Noise suppressor
S
Circuitry
Push - Pull
Battery Type
LR03, AM4
Supply Voltage
1.5 V
Battery Current
9.9 mA
Battery Life
96 h
Dimensions
6.3 × 4.2 × 1.4 cm
Weight (excluding battery)
24 g
(Typical value)
TECHNICAL DATA
(
According to IEC Standard Pub. 60118-0 1983
)
Reference Test Frequency
1600 Hz
OSPL
90
138 dB
500 Hz
135 dB
Peak
142 dB
Full on Acoustic Gain
74 dB
Equivalent Input Noise Level
19 dB
Total Harmonic
Distortion
500 Hz
16%
800 Hz
23%
1600 Hz
3%
Induction Coil Sensitivity
113 dB at 1 mA / m
Operating Switch
- S - / M-MT
Output Limiting Control
MOP, range 20 dB (continuous)
Tone Control
TONE H, TONE L (continuous)
Noise suppressor
S
Circuitry
Push - Pull
Battery Type
LR03, AM4
Supply Voltage
1.5 V
Battery Current
5.9 mA
Battery Life
160 h
Dimensions
6.3 × 4.2 × 1.4 cm
Weight (excluding battery)
24 g
(Typical value)
2. When locking the control for your comfortable
level, open the housing case, slide the lock
key downward.
Do not turn the control at the "LOCK" posi-
tion, because this may break.
When setting the Lock key to "FREE", it will be released.