88
Commands
Auto 1 store example
(Frequency weighting: C, Time weighting: FAST, Level range: 40 to 130 dB,
Measurement time: 5 min, File name: AU1_0001, Sampling cycle: 100 msec)
Power ON, CF card inserted
CDV? (Confi rm inserting a CF card)
Settings
- WGT1_# (Main:
C)
- TMC0_# (Main:
F)
- RNG5
(40 to 130 dB)
- MTI5_1 (5
m)
- SMD1
(AUTO1)
- SNS0001 (File
name)
- PLP100_# (100
msec)
Establish the above settings to prepare the Auto1 measurement.
STO1
(Start store operation)
SRT0 (Stop
measurement)
Data are stored on the CF card as a text fi le.
To copy the data to the computer, access the card recognized by the computer
as “removable media”.