Turn-on precaution
If the hearing aid is turned on again immediately after turning it off,
there may be no sound, but this is not a defect. In such a case,
turn the power off (open battery holder) and wait 10 seconds.
Then turn the power on again (close battery holder).
Removing Hearing Aid
Carefully pull the wire to remove the hearing aid. When acoustic
feedback occurs, open the battery holder (power off). Acoustic
feedback may stop. Then pull the wire first sideways and then
up.
Adjusting the Controls
This hearing aid is a digital type. Any adjustments must be done
at the dealer. For details, please contact the dealer.
Cleaning your Hearing Aid
We recommend that you clean
your hearing aid twice a week
(more or less needed) to remove
any earwax that may build up as
a result of regular use. Too much
earwax build up can prevent your
hearing aid from working.
Hold the hearing aid as the il-
lustration, brush off the earwax
of the sound outlet.
TECHNICAL DATA
(
According to ANSI standard S3.22 2003
)
Maximum-OSPL90
109 dB (3900 Hz)
HFA-OSPL90
106 dB
HFA-full-on Acoustic Gain
37 dB
Reference Test Gain
29 dB
Frequency Range
200 Hz to 5000 Hz
Equivalent Input Noise Level
27 dB
Total Harmonic Distortion 500 Hz: 2% 800 Hz: 2% 1600 Hz: 2%
Operating Switch
Case Switch
Output Limiting Control
MOL
Gain Control
GAIN (range: 67 dB)
Battery Type/Supply Voltage 10/1.3 V
Battery Current
0.74 mA
Battery Life
Approx. 100 hours
(Typical value)
TECHNICAL DATA
(
According to IEC standard Pub. 60118-0-1983
Amendment 1-1994
)
Reference Test Frequency 1600 Hz
OSPL
90
104 dB
500 Hz
112 dB
Peak
122 dB
Full-on Acoustic Gain
45 dB
Equivalent Input Noise Level
25 dB
Total Harmonic Distortion 500 Hz: 3% 800 Hz: 3% 1600 Hz: 2%
Operating Switch
Case Switch
Output Limiting Control
MOL
Gain Control
GAIN (range: 67 dB)
Battery Type/Supply Voltage PR536/1.3 V
Battery Current
0.74 mA
Battery Life
Approx. 100 hours
(Typical value)
Carefully pull on
pull-out wire
No. 51340 06-11
110
0dB=20µPa
1600 2500
100
90
80
70
60
100
200
500
1000
2000
5000
10000
Frequency (Hz)
O
ut
pu
tS
P
L
(d
B
)
Input SPL 60dB
60
1600 2500
50
40
30
20
10
100
200
500
1000
2000
5000
10000
Input SPL 50dB
(0dB=20µPa)
Frequency (Hz)
A
co
us
tic
G
ai
n
(d
B
)
130
0dB=20µPa
1600 2500
120
110
100
90
80
100
200
500
1000
2000
5000
10000
Frequency (Hz)
O
ut
pu
tS
P
L
(d
B
)
Input SPL 90dB
110
0dB=20µPa
1600 2500
100
90
80
70
60
100
200
500
1000
2000
5000
10000
Frequency (Hz)
O
ut
pu
tS
P
L
(d
B
)
Input SPL 60dB
130
0dB=20µPa
1600 2500
120
110
100
90
80
100
200
500
1000
2000
5000
10000
Frequency (Hz)
O
ut
pu
tS
P
L
(d
B
)
Input SPL 90dB
60
1600 2500
50
40
30
20
10
100
200
500
1000
2000
5000
10000
Input SPL 50dB
(0dB=20µPa)
Frequency (Hz)
A
co
us
tic
G
ai
n
(d
B
)
Full-on gain curve
Frequency response curve
OSPL90 curve
Full-on acoustic gain frequency
response curve
Basic frequency response
curve
OSPL
90