FLIGHT: Set the amount of material that is in-flight. This is used to turn the output off before the
target it reached so it does not overshoot.
LOCK: Set if the target and flight values will be locked from editing in the OPER menu.
SOURCE: Select the weight source for the setpoint to use. Options are:
– GROSS uses gross weight only
– NET uses net weight only
16.6
Status Based Setpoint Types
The following setpoint types are all based on the status of the instrument.
NONE: Setpoint is always inactive.
ON: Setpoint is always active. This type of setpoint is useful to show that the instrument is running.
Centre of Zero (COZ): Setpoint is active when the gross weight is within COZ bounds.
ZERO: Setpoint is active when the gross weight is within the Zero Band setting.
NET: Setpoint is active when the NET annunciator is lit.
MOTION: Setpoint is active when the MOTION annunciator is lit.
ERROR: Setpoint is active when the instrument detects any error condition signified by the display of
Exxxxx on the display.
BUZZER: Setpoint is active when the buzzer beeps.
16.7
Example 1 (filling a bag)
Type = Over, Logic = High, Source = Gross, Target = 2000kg, Inflight = 50 kg, Hysteresis = 5 kg. Initial
weight = 0 kg.
Trip point = target - inflight = 2000 - 50 = 1950 kg.
The output will switch ON at a weight over 1950 kg and switch OFF again at a weight under 1945 kg.
If the type is changed to weigh in the output will be ON at 0 kg, switch OFF at a weight over 1950 kg and
switch ON again at a weight under 1945 kg.
16.8
Example 2 (loss in weight system)
Type = Under, Logic = High, Source = Net, Target = -100kg, Inflight = 5 kg, Hysteresis = 1 kg. Initial
weight = 0 kg.
Trip point = inflight = -100 + 5 = -95 kg.
The output will switch ON at a weight under -95 kg and switch OFF again at a weight over -94 kg.
68
C500-600-2.0.3
Содержание C500 series
Страница 1: ...C500 Series Digital Indicator Reference Manual ...
Страница 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Страница 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Страница 120: ...Select C500 and click create Enter the module name description and IP address then click change 115 C500 600 2 0 3 ...
Страница 123: ...118 C500 600 2 0 3 ...
Страница 182: ...177 C500 600 2 0 3 ...