6 Operations and Functions

 

6-2. Fault alarm activation 

 

 

- 34 - 

 

<Response to Alarm> 

An oxygen concentration value exceeds the alarm setpoint 

When an alarm is triggered, take actions in accordance with your management rules of alarm. 
Normally, take the following actions. 
 

 

Check the reading of the detector. 

 

NOTE 

If a gas leak is momentary, the reading may already have dropped when checking it. In addition, when the 
alarm is triggered by noise or other incidental conditions other than a gas, the reading may have already 
dropped. 
 
 

 

Based on your management rules of gas alarm, no one can be allowed to access the monitored 
zone to ensure safety. 

 

If the gas concentration display continues to be displayed, close the main valve of the gas, and then 
check that the gas concentration reading dropped. 

 

Access the gas leak point, equipped with a protective gear to avoid dangers caused by possibly 
remaining gases, and check whether gases remain or not by using a portable gas detector. 

 

Check that the point is free from dangers, and take actions to fix the gas leak. 

 
 

6-2. Fault alarm activation 

 
A fault alarm is triggered when the detector detects abnormalities. After a fault alarm is triggered, the fault 
lamp (yellow) lights up and an error message is displayed on the LED. Determine the causes and take 
appropriate actions. 
After the detector is successfully returned from the fault, it restarts with the process normally performed right 
after it is turned on (initial clear). 
If the detector has problems and is repeatedly malfunctioning, contact RIKEN KEIKI immediately. 
 

<Display Operation> 

Fault Detail Display 

Display a message indicating the fault detail. 
 

Fault Lamp (FAULT: Yellow) 

This lights up when a failure occurs. 
 

NOTE 

For information on malfunctions (error messages), see "9. Troubleshooting". 
 
 
 

Содержание SD-1DOX-AS

Страница 1: ...PT2E 2625 Smart Transmitter Gas Detector Head SD 1DOX AS TYPE H Operating Manual PT2E 179 PT2 179...

Страница 2: ...addition it may be fluctuated by environmental temperature humidity etc changes in the installation site 2 The alarm must be set within a range where the performance of the detector can be ensured The...

Страница 3: ...o wire 21 4 7 How to tube 25 5 How to Operate 26 5 1 Preparation for start up 26 5 2 Basic operating procedures 26 5 3 How to start the gas detector 27 5 4 Modes 28 5 5 User mode 29 5 6 Flow Rate Adju...

Страница 4: ...s concentration is over the setting value The detector is a safety unit not an analyzer or densitometer which performs quantitative qualitative analysis measurement for gases Please fully understand t...

Страница 5: ......

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ......

Страница 10: ......

Страница 11: ...standard accessories Main Unit Standard Accessories Operating manual one Dedicated handling lever one Dedicated control key The supplied quantity depends on the number of units to be delivered 1 to 10...

Страница 12: ......

Страница 13: ......

Страница 14: ...o switch menus or change a value DOWN 5 Power lamp Power lamp Detection mode It lights up in green Maintenance mode It blinks in green 6 Alarm lamp Alarm lamp It lights up in red when the alarm setpoi...

Страница 15: ......

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ...ning is impossible the following protective measures can be taken Protection against lightning Take appropriate measures in accordance with the importance of the facilities and the environment Provide...

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ......

Страница 29: ......

Страница 30: ......

Страница 31: ......

Страница 32: ......

Страница 33: ......

Страница 34: ......

Страница 35: ...d if not needed Display Operation Oxygen Concentration Display In case of over the detection range Over Scale is displayed on the LED Power Indicator Lamp POWER Green During operation this lights up c...

Страница 36: ...cess the gas leak point equipped with a protective gear to avoid dangers caused by possibly remaining gases and check whether gases remain or not by using a portable gas detector Check that the point...

Страница 37: ......

Страница 38: ......

Страница 39: ...as a safety unit Maintenance item Maintenance content Daily mainte nance Monthly mainte nance Regular mainte nance Power Supply Check Check that the power lamp lights up Concentration Display Check C...

Страница 40: ...r to find abnormalities Checks the flow rate by using an external flow meter to verify the correctness of the flow rate indicator on the gas detector If the flow rate is incorrect performs the flow ra...

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ......

Страница 45: ......

Страница 46: ......

Страница 47: ......

Страница 48: ......

Страница 49: ......

Страница 50: ......

Страница 51: ......

Страница 52: ......

Страница 53: ......

Страница 54: ......

Страница 55: ...vide the rated voltage Take measures such as checking or adding the UPS power supply line filter and insulation transformer Cable abnormalities open circuit not connected short circuit Check the wirin...

Страница 56: ...ently take appropriate measures to eliminate the noise Sudden change in the environment When the environment temperature etc changes suddenly the detector cannot adjust to it and is affected by it In...

Страница 57: ...ition Less than 1vol against alarm preset point Alarm delay time 1 under an identical condition By anoxia alarm Alarm preset point 18vol Less than 5sec When introducing 10 11vol gas Without piping del...

Страница 58: ...let Bs union for Cu pipe for O D 8 1 Compressed air inlet BS union for Cu pipe for O D 8 1 Initial clear Approx 25sec Operating temperature 10 40 non rapidly vary Operating humidity Less than 95 RH no...

Страница 59: ...analog transmission digital transmission HART communication or Three wire analog transmission in common with power supply power supply signal common Specification of Transmission DC4 20mA linear load...

Страница 60: ...o wire analog transmission digital transmission HART communication or Three wire analog transmission in common with power supply power supply signal common Specification of Transmission DC4 20mA linea...

Страница 61: ...10 Product Specifications 10 1 List of specifications 59 Outline Drawings Adapter A NPT1 2...

Страница 62: ...1a Non exciting at normal exciting at alarm or exciting at normal non exciting at alarm Contact capacity AC250V 0 5A DC30V 0 5A resistive load Transmission scheme Two wire analog transmission digital...

Страница 63: ...ver Transmission scheme Two wire analog transmission digital transmission HART communication or Three wire analog transmission in common with power supply power supply signal common Specification of T...

Страница 64: ...o recover Transmission scheme Two wire analog transmission digital transmission HART communication or Three wire analog transmission in common with power supply power supply signal common Specificatio...

Страница 65: ...10 Product Specifications 10 1 List of specifications 63 Outline Drawings Cable gland G3 4...

Страница 66: ......

Страница 67: ...power This is a function to prevent triggering alarm during that time Full scale Maximum value of the detection range Calibration Adjusts the readings to the calibration gas concentration value by usi...

Страница 68: ...0 First issue 3 22 2018 1 Declaration of Conformity 12 23 2019 2 Declaration of Conformity 4 1 2020 3 Safety information Declaration of Conformity 8 1 2020 4 Safety information Declaration of Conformi...

Страница 69: ......

Отзывы: