6
・
Connect gas sampling probe and gas sampling hose (spiral) to
the instrument (gas inlet) firmly.
Otherwise, correct measuring can not be performed.
・
Keep the gas sampling probe (with absorbent cotton) on the
instrument all the time during gas detection.
・
Without filter, the dust, water and oil are sucked in to the
instrument. These material might be a cause of breakdown.
2-2.How to start
By pressing ON/OFF switch for 3 seconds, the power turns on and starts gas detection,
through following procedure, battery voltage check, tube/absorbent cotton check and
sensor connection check. The starting flow is as follows.
Press ON/OFF switch for 3 seconds.
CO2 0-10000 ppm
(Indicate measuring gas and range.)
Approx. 5 sec. At pump connection failure.(P.7 ①)
BATT.:E ---|-- F
(Indicate battery capacity sample.)
Approx. 5 sec. At battery voltage shortage.(P.7 ②)
PREPARATION
(Prepare before detection.)
Approx. 5 sec.
PROBE/FILTER OK?
(・Connect gas sampling probe.)
Approx. 5 sec. If OK, press ENTER switch (・Replace absorbent cotton.)
SENSOR CHECK
(Self diagnosis sensor check.)
At sensor connection failure.(P.7 ③)
SENSOR CHECK OK
Approx. 5 sec.
WARMING UP 30sec
(Warming up 30 sec.)
30 sec.
Caution
!
When power on, check the buzzer
and pump sound.
*
Note