17
4-6 Suitable cables
CAUTION
・
Use junction box if you are using a wire that cannot go through the cable port of
RI-2000R.
4-6-1 Power cable
Please use relevant cables to CVVS 1.25 to 2.0 mm
2
.
4-6-2 4-20mA signal/ contact signal line
Please use signal lines relevant to CVVS 1.25 to 2.0mm
2
.
4-6-3 Terminal block specification
【
Terminal block specification
】
・
Power input : 3P
・
Each signal output : 8P
・
Terminal size : M4
・
Tightening torque : 1.2N
・
m
4-7 Terminal block diagram
CAUTION
・
After having wired the terminal block, please attach a cover by all means not to get
shocked.
Power input
AC200/220V
“Warning” alarm
output contact
“Alarm” alarm
output contact
“Trouble” alarm
output contact
4-20mA
Signal output