6 Maintenance

 

 

6-1. Maintenance intervals and items 

 

 

- 21 - 

 

 
 
 
 
 

Maintenance 

 
 
 
 
The detector is an important instrument for the purpose of safety. 
To maintain the performance of the detector and improve the reliability of safety, perform a regular 
maintenance. 
Continuing to use the detector without performing maintenance will compromise the sensitivity of the gas 
sensor, thus resulting in inaccurate detection. 
 

6-1. Maintenance intervals and items 

 

 

Daily maintenance: Perform maintenance before beginning to work. 

 

Monthly maintenance: Perform maintenance on the alarm circuit (alarm test) once a month. 

 

Regular maintenance: Perform maintenance once or more for every six months to maintain the 
performance as a safety unit. 

 

Maintenance 

item 

Maintenance content 

Daily 

mainte-

nance 

Monthly 

mainte- 

nance 

Regular 

mainte- 

nance 

Power Supply 
Check* 

Check that the power lamp lights up. 

 

 

 

Status display 
Check* 

Check that the fault lamp does not light up. 

 

 

 

Concentration 
Display Check* 

Check that the concentration display value is zero (or 
20.9 vol% on the oxygen meter). When the reading is 
incorrect, perform the zero adjustment (or span 
adjustment) after ensuring that no interference gases 
exist around it. 

 

 

 

Alarm Test* 

Inspect the alarm circuit by using the alarm test function. 

 

 

 

Span 
Adjustment* 

Perform the span adjustment by using the calibration 
gas. 

 

 

 

Gas Alarm 
Check* 

Check the gas alarm by using the calibration gas. 

 

 

 

* Check and adjustment are performed at the indicator/alarm unit side. See the operating manual of the 

indicator/alarm unit for details. 

 
 

NOTE 

 

Before starting alarm test or span adjustment, be sure to provide a notification to the related 
sections and block external signals. 

 
 
 
 
 
 
 

Содержание GD-10X Series

Страница 1: ...PT2E 1822 Fixed type Oxygen Detector Head GD 1OX Series GD 1OX GD 1DOX Operating Manual PT2 182 PT2 183...

Страница 2: ...pressure other than atmospheric pressure a calibration is required under that pressure The maximum allowable pressure is 110 KPa The detector cannot be used under reduced pressure 3 The gas sensor use...

Страница 3: ...tem designing 10 4 4 How to install 11 4 5 Installation procedure 13 4 6 How to wire 15 4 7 Cable connection example 17 4 8 How to tube 19 5 How to Operate 20 5 1 Preparation for start up 20 5 2 Basic...

Страница 4: ...that detects oxygen and performs the alarm activation when the gas concentration is over the setting value The detector is a safety unit not an analyzer or densitometer which performs quantitative qu...

Страница 5: ......

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ......

Страница 10: ......

Страница 11: ......

Страница 12: ......

Страница 13: ......

Страница 14: ...d by RIKEN KEIKI BARRIERTA JFE 552 manufactured by NOK KLUBER If you can not prepare the specified grease use one that meets the following requirements 1 Material does not harden due to deterioration...

Страница 15: ......

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ...tenance content Daily mainte nance Monthly mainte nance Regular mainte nance Power Supply Check Check that the power lamp lights up Status display Check Check that the fault lamp does not light up Con...

Страница 24: ...l on the oxygen deficiency meter by using the zero gas Performs the zero adjustment fresh air adjustment if the reading is incorrect Alarm Test Inspects the alarm circuit by using the alarm test funct...

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ......

Страница 29: ......

Страница 30: ......

Страница 31: ...s to indicator alarm unit as well See the operating manual of the indicator alarm unit for details Symptom Display FAULT Causes Actions Sensor error appears on the indicator side The wiring is not cor...

Страница 32: ...H 88 D mm projection portions excluded Weight Approx 2 5kg Color Munsell 7 5BG5 2 Suction type Model GD 1DOX Detection principle Galvanic cell method Detection gas O2 Detection method Suction method...

Страница 33: ......

Страница 34: ...used to express the percentage of a specific substance or gas in a volume of solution Atmosphere An atmosphere with a temperature within the range of 10 to 40 C and a humidity of 95 RH or less at one...

Страница 35: ...Manual Log Rev Amendment Issue data 0 First issue 2018 2 16 1 Completely revised GD 1DOX model added 2018 12 19 2 P11 Corrected maintenance space 2020 6 29...

Отзывы: