Symptom/display
Cause
Action
The power cannot
be turned on.
The power switch is
turned off.
Turn the power switch ON.
Cable abnormality
(break, unconnected,
short)
Check the wiring for this product and for related devices.
The AC adapter is
defective.
The AC adapter must be replaced.
Contact Riken Keiki.
The AC adapter is not
correctly connected.
Connect the AC adapter correctly.
Abnormal operation
Effects of radio waves,
etc.
Turn off the power and restart.
If similar symptoms recur frequently, take appropriate
measures to address the noise source.