4. PLC Communication Functions
4-3. Communication settings
- 50 -
4-3. Communication settings
GD-70D-EA Settings can be configured through the Network screen displayed by logging in as an
authorized user using the Web function, or through the front panel keys. Note that when using the front
panel the configuration is restricted.
See Chapter 2, “Web Functions” for details of Web functions. See 4-6, “Specifications for communication
setting through GD-70D-EA main unit operation” for details of specifications for communication setting
through the front panel.
Item
Description
Remarks
PLC PLC
Type
Choose the communication type of the
PLC.
FINS settings in this table are used if
FINS is selected, or MC settings in this
table are used if MC is selected.
Not Used: PLC is not used
(default)
FINS: OMRON
CJ/CS
Series
MC:
MELSEC Q Series
Type of Basic Data
Type of Basic Data
Small: Small
Large: Large
Use Optional Data
With or without use optional data
Yes: With
use
No: Without
use
(default)
Interval
Min
Minimum transmission interval.
Transmission interval in abnormal
measuring condition (in alarm or
maintenance state) with data changing.
250 to 10000 msec
Minimum time default:
1000 msec
(With variation in
concentration)
Maximum time default:
3000 msec
(Without variation
in concentration)
Max
Maximum transmission interval.
Transmission interval in normal
measuring condition without change in
data.
Timeout
Timeout period of receiving reply from
the PLC.
Stops waiting for a reply from the PLC
following transmission to the PLC if no
reply has been received when
predefined waiting time elapsed.
Transmission occurs again when the
length of time set in Interval elapsed.
1 to 10 sec
Default: 5 sec
FINS
IP Address
IP address of the PLC
Port
UDP port number
Default:
9600
Detector
Network
Address
Network address of GD-70D-EA
(Setting value for FINS specification)
Node
Address
Node address of GD-70D-EA
(Setting value for FINS specification)
Unit
Number
Unit number of GD-70D-EA
(Setting value for FINS specification)
PLC
Network
Address
Network address of the PLC
(Setting value for FINS specification)
Node
Address
Node address of the PLC
(Setting value for FINS specification)
Unit
Number
Unit number of the PLC
(Setting value for FINS specification)
Detector
Write
Memory
Address
Memory address used when writing to
PLC
Area Type
Memory area used when writing to PLC
Detector
Read
Memory
Address
Memory address used when reading
from PLC
Area
Type
Memory area used when reading from
PLC
Содержание 70D Series
Страница 1: ...PT2E 1582 Gas detector head 70D Series Ethernet Model Communication Function Manual...
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 11: ......
Страница 13: ......
Страница 14: ......
Страница 15: ......
Страница 16: ......
Страница 17: ......
Страница 18: ......
Страница 19: ......
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 36: ......
Страница 45: ......
Страница 46: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 52: ......
Страница 56: ......
Страница 57: ......
Страница 60: ......
Страница 61: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......