14
4−5.
Grounding
Make grounding with terminal
Be sure to make grounding before putting power ON.
Be sure to make grounding for safety purpose and to keep stable operation of
this unit.
Do not connect earth wire with gas pipe. Make grounding corresponding to
D-class grounding (Grounding resistance : below 100 ohms).
WARNING
Terminal